Table 4.

Accuracy of Left Ventricular EF Estimates Relative to Ground Truth EF of 0.403

3‐Axis Eqn. 5 LA‐Area‐Length Eqn. 6 Bi‐Plane LA Eqn. 7 Dodge Eqn. (8) Teichholz Eqn. (9)
First slice EF 0.308 4.19 0.407
First slice EF error, % −23.6 4.0 1.1
Mean EF (n=12 or 6*) 0.271 0.449 0.470* 0.507* 0.361
EF mean error, % −32.8 11.4 16.7 26.0 10.3
EF SD error, % 10.5 20.0 10.0 11.1 23.0
EF error range, % −49.9 to −14.6 −20.71 to 42.1 5.5 to 29.0 13.5 to 39.7 −52.5 to 36.3

When using Equations 5, 6, and 9, the mean and SD were computed over 12 independent EF measurements (1 for each long‐axis [LA] slice). The error metrics for Equations 7 and 8 are based on 6 pairs of orthogonal long‐axis slices, as indicated by the asterisks. EF indicates ejection fraction.