Table 2.
All laryngoscopes | Pentax-Airway Scope | King VISION | McGrath Mac | Macintosh | P value | |
All operators, n | 287 | 82 | 59 | 82 | 64 | |
First-pass intubation success | 199 (69) | 64 (78)* | 34 (58) | 64 (78)† | 37 (58) | 0.004 |
Non-expert operators | 156 | 46 | 32 | 37 | 41 | |
First-pass intubation success | 104 (67) | 40 (87)‡ | 16 (50) | 29 (78)§ | 19 (46) | 0.00004 |
Expert operators | 131 | 36 | 27 | 45 | 23 | |
First-pass intubation success | 95 (73) | 24 (67) | 18 (67) | 35 (78) | 18 (78) | 0.556 |
Values are numbers (%) of observations; post hoc analyses were performed using Tukey-Kramer test for paired comparisons of four laryngoscopes.
*vs King VISION p=0.043; vs Macintosh p=0.039.
†vs King VISION p=0.043; vs Macintosh p=0.039.
‡vs King VISION p=0.002; vs Macintosh p<0.001.
§vs King VISION p=0.043; vs Macintosh p=0.009.