Table 4:

Temperature rise in the gel at the tip of ipsilateral and contralateral leads for RF exposures of Table 3.

Temperature rise [℃] 1.5 T (64 MHz) 3 T (127 MHz)
Ipsilateral Contralateral Ipsilateral Contralateral
Wire 1.4 3.6 2.3 3.3
Medtronic Lead 3389 0.8 2.3 0.2 2.0
Medtronic Lead 3387 0.3 1.2 0.2 1.0
Wire, modified trajectory 1.4 0.6 2.3 0.5