1. Preparation and communication
-
a
Make a list of patients with affected devices with phone number, address, type of device and date of implant
-
b
Exclude patients from the list that have known to have expired, or who have undergone cardiac transplant, or device explantation due to infection or other reasons
|
2. Out-patient follow up
-
a
Add “note” and check “highlight at every follow up” (on the device)
-
b
Interrogation including battery longevity
-
c
Assess pacing dependency (ICD and CRT-D) and percentage pacing in ICDs
-
d
Ensure vibratory alert is ON for ERI
-
e
Demonstrate and familiarize the patient to the Vibratory Alert
-
f
Assess present ICD indication in patients with primary prevention ICD indication
-
g
Request for a 3-month follow-up in those without remote monitoring
|
| 3. Remote monitoring – Merlin.net for all/selected patients (see Table 3) |
| 4. Consider prophylactic device replacement for highly selected patients (see Table 4) |
| 5. Notification to IHRS if a patient with premature battery depletion with or without any adverse event/s is seen |