Skip to main content
. 2019 Mar 30;21(4):155–158. doi: 10.1089/dia.2018.0359

Table 1.

Urgent Low Soon Activations and Mean (Standard Deviation) Continuous Glucose Monitoring–Derived Glycemic Metrics of G5 Users Who Transitioned to G6 with the Urgent Low Soon Alert Enabled

  Low threshold alert setting of 70 mg/dL (n = 658) Low threshold alert setting of 80 mg/dL (n = 766)
G5 G6 P-value G5 G6 P-value
ULS activations/day 0.9 (0.8) 0.6 (0.6)
Mean glucose (mg/dL) 165.4 (33.0) 163.0 (31.4) 0.18 174.2 (30.3) 172.2 (29.8) 0.19
Time <54 mg/dL (%) 1.0 (1.5) 0.6 (1.0) <0.001 0.6 (1.1) 0.4 (0.8) <0.001
Time ≤55 mg/dL (%) 1.2 (1.7) 0.8 (1.2) <0.001 0.7 (1.3) 0.5 (0.9) <0.001
Time <70 mg/dL (%) 3.7 (3.5) 3.0 (3.0) <0.001 2.3 (2.8) 2.0 (2.5) 0.006
TIR (70–180 mg/dL) (%) 61.1 (17.9) 63.0 (18.1) 0.05 57.5 (17.5) 58.7 (18.1) 0.18
Time >180 mg/dL (%) 35.2 (19.1) 34.0 (19.1) 0.23 40.2 (18.4) 39.4 (18.9) 0.37
Time >250 mg/dL (%) 12.6 (12.1) 11.1 (11.2) 0.02 14.4 (11.8) 13.0 (11.5) 0.02

G5, fifth-generation; G6, sixth-generation; TIR, time in range; ULS, “Urgent Low Soon”.