Skip to main content
. 2019 Mar 30;21(4):201–208. doi: 10.1089/dia.2018.0376

FIG. 4.

FIG. 4.

Basal rate accuracy of 1-h windows of all insulin pump systems at 1.0 U/h. For each system, mean (green diamonds), median with first and third quartile (black boxes), and 95% range (gray boxes) are shown (n = 648 values). Red lines and red dashed lines indicate target ±5% and target ±15%, respectively.