Figure 1.

PED clinical prototype. (A) Schematic for PED is shown. The electroceutical dressing consists of an interdigitated silver ink pattern on silk fabric; the dressing is powered by a 6 V battery (B) (Model No. 36-3025-2-ND) with a 10 kΩ resistor (R) that limits the current in the active circuit. GND, ground; SW, switch (Model No. 563-1151-ND). (B) Fully assembled PED-10 for clinical use. PED, patterned electroceutical dressing; PED-10, dressing with 10 kΩ ballast resistor.