Skip to main content
. 2019 Apr 23;9:6370. doi: 10.1038/s41598-019-42783-w

Table 1.

Characteristics of the investigated dialysersa.

Dialyser brand name Membrane Membrane polymer Treatment modality Surface area (m2) Fibre ID/weight (µm) ± 2% Water permeability (mL/m2 * h * mmHg) ± 5%
Theranova 500 MCO PES/PVP HD mode 2.0 180/35 636
Polyflux 210 H High flux PES/PVP HD and HDF modes 2.1 215/50 312
FX CorDiax 800 High flux PSu/PVP HDF mode 2.0 200/45 276

HD, haemodialysis; HDF, haemodiafiltration; MCO, medium cut-off PES, polyethersulfone; PSu, polysulfone; PVP, polyvinylpyrrolidone.

aSurface areas stated are as listed in published data sheets, fibre dimensions measured for 20 fibres in one filter and water permeability (as flow of purified water at a given transmembrane pressure per membrane surface at 37 ± 1 °C) measured for 6 filters.