Skip to main content
. 2019 Apr 2;24(7):1295. doi: 10.3390/molecules24071295

Table 1.

Statistics of quality control (QC) levels from the qualification run for tozadenant in rat plasma.

Run Number Statistics QC Low
(15.04 ng/mL)
QC Medium
(165.46 ng/mL)
QC High
(1820 ng/mL)
1 Mean
n
% Acc
% CV
15.94
3
106.01
6.81
171.91
3
103.90
1.32
1802.27
3
99.03
4.62
2 Mean
n
% Acc
% CV
14.95
3
99.38
6.19
157.76
3
95.34
1.13
1684.91
3
92.58
0.24
3 Mean
n
% Acc
% CV
15.32
3
101.88
4.34
163.48
3
98.80
1.78
1807.70
3
99.32
0.51
Inter-run Mean
n
% Acc
% CV
15.40
9
102.42
5.84
164.38
9
99.35
3.95
1764.96
9
96.98
4.15

% Acc = inter-run accuracy, % CV = precision.