Skip to main content
. 2019 Apr 10;2019:5643848. doi: 10.1155/2019/5643848

Table 2.

Validation parameters of the methods used for the determination of microelements.

Microelement/method Dilution factor Recovery of the CRM (%) Detection limit Accuracy (%) Precision (coefficient of variation) (%)
Cu/FAAS 25 95 0.158 ppm 2.2 5.1
Zn/FAAS 250 101 0.013 ppm 1.3 2.1
Mn/ET AAS 125 98 0.084 ppb 1.2 2.1
Se/ETAAS 25 99 1.74 ppb 0.4 1.9

FAAS: flame atomic absorption spectrometry; ETAAS: electrothermal atomic absorption spectrometry; ppm: µg/g; ppb: ng/g.