Figure 2.

Figure 2.

Left panel: experimental DMF cartridge for performing enzymatic assays by fluorometry. There are 12 sample inputs (right edge), 8 reagent inputs (top and bottom edge) plus additional reservoirs for stop buffer and waste. Individual droplets are dispensed, mixed and manipulated entirely by software. Right panel: results of enzymatic assays for the enzymes deficient in Fabry (GLA) disease and Pompe (GAA) disease by DMF using a 1 h incubation time. Results for known affected patients (GLA: n=6; GAA: n=7) are compared with presumed unaffected controls (n=105) and show no overlap. Data modified and reprinted with permission from the American Association of Clinical Chemistry [12].