Published in final edited form as: Acad Radiol. 2017 Mar 2;24(7):876–890. doi: 10.1016/j.acra.2016.12.017

Table 4.

Number of successfully interpreted cases for each diagnostic task - at every dose-reconstruction configuration - using binary reader agreement rules. For positive cases, a majority of readers had to identify all lesions seen by a majority of readers using the routine dose level; for negative cases, the majority of readers made no non-lesion localizations (i.e., no false positive markings). Numbers in red indicate failure to meet preset thresholds for successful dose-reconstruction configurations. The number of cases with essential lesions reflects cases with target lesions identified by 2 or more readers at routine dose level.

Contrast-enhanced abdominal CT for Hepatic Metastasis
Dose Level by AEC setting (QRM) Reconstruction Kernel (FBP/IR) # of readers with successful interpretation Successful Interpretations
All Cases
Successful Interpretations
Cases with (n=21): Cases without (n=23) essential Lesions
0/3 1/3 2/3 3/3
200 B30 0 3 9 32 41 20:21
I30 0 3 9 32 41 19:22
160 B30 0 4 12 28 40 18:22
I30 0 4 10 30 40 18:22
120 B30 2 4 5 33 38 16:22
I30 1 4 8 31 39 18:21
100 B30 3 6 7 28 35 17:18
I30 2 4 9 29 38 17:21
80 B30 3 6 10 25 35 14:21
I30 4 4 8 28 36 14:22
60 B30 4 4 10 26 36 14:22
I30 7 4 8 25 33 14:19
Unenhanced chest CT for Pulmonary Nodule detection
Dose Level by AEC setting (QRM)) Reconstruction Kernel (FBP/IR) # of readers with successful interpretation Successful Interpretations
All Cases
Successful Interpretations
Cases with (n=19): Cases without (n=25) essential Lesions
0/3 1/3 2/3 3/3
70 B50 1 3 12 28 40 19:21
I50 1 1 10 32 42 19:23
30 B50 0 3 9 32 41 17:24
I50 1 3 7 33 40 17:23
10 B50 0 4 8 32 40 18:22
I50 1 2 5 36 41 17:24
5 B50 1 2 7 34 41 18:23
I50 1 1 6 36 42 19:23
2.5 B50 2 6 4 32 36 13:23
I50 2 3 8 31 39 16:23
Unenhanced spiral head CT to detect Intracranial Findings Causing Neurologic Deficit
Dose Level by effective mAs Reconstruction Kernel (FBP/IR) # of readers with successful interpretation Successful Interpretations
All Cases
Successful Interpretations
Cases with (n=18): Cases without (n=25) essential Lesions
0/3 1/3 2/3 3/3
250 H40(FBP) 0 2 2 39 41 16:25
J40 (IR) 0 1 7 35 42 17:25
200 H40(FBP) 0 2 4 37 41 17:24
J40 (IR) 0 2 7 34 41 17:24
150 H40(FBP) 0 1 6 36 42 17:25
J40 (IR) 1 0 4 38 42 17:25
100 H40(FBP) 0 0 10 33 43 18:25
J40 (IR) 0 2 6 35 41 16:25
50 H40(FBP) 1 1 9 32 41 17:24
J40 (IR) 0 1 6 36 42 17:25
25 H40(FBP) 2 3 8 30 38 14:24
J40 (IR) 2 0 11 30 41 16:25

AEC = automatic exposure control setting, which adjusts the modulation of the CT system tube current; QRM = quality reference milliampere-seconds (mAs), the AEC setting that will produce the image quality at a similar effective mAs for a hypothetically normal-sized patient; FBP = filtered back projection; IR = iterative reconstruction