. 2019 Apr 25;9:6557. doi: 10.1038/s41598-019-42968-3

Table 2.

Comparison of prediction errors for each formula using two devices.

Arithmetic prediction error within (%) Formula Optical Biometer P value
ARGOS IOLMaster 700
Over entire AL range (n = 106)
±0.50 D Barrett Universal II 66.0 62.3 0.481
Haigis 68.9 68.9 1
Hoffer Q 58.5 56.6 0.815
SRK/T 63.2 62.3 1
±1.00 D Barrett Universal II 92.5 91.5 1
Haigis 93.4 93.4 1
Hoffer Q 92.5 88.7 0.289
SRK/T 92.5 91.5 1
Medium AL group (n = 76)
±0.50 D Barrett Universal II 64.5 61.8 0.791
Haigis 67.1 71.1 0.607
Hoffer Q 67.1 67.1 1
SRK/T 61.8 68.4 0.277
±1.00 D Barrett Universal II 92.1 90.8 1
Haigis 93.4 93.4 1
Hoffer Q 93.4 93.4 1
SRK/T 92.1 89.5 0.625
Long AL group (n = 30)
±0.50 D Barrett Universal II 70.0 63.3 0.625
Haigis 73.3 63.3 0.250
Hoffer Q 36.7 30.0 0.625
SRK/T 66.7 46.7 0.070
±1.00 D Barrett Universal II 93.3 93.3 1
Haigis 93.3 93.3 1
Hoffer Q 90.0 77.7 0.125
SRK/T 93.3 96.7 1