Table 1.

Border sharpness, image quality, and scan time for whole-heart MRA for SENSE x2 versus CS x6 (n=15).

SENSE x2
CS x6
p-value
Border sharpness Image quality Border sharpness Image quality Border sharpness Image quality



LLPV 0.69 ± 0.13 4.17 ± 0.75 0.74 ± 0.17 3.27 ± 0.64 0.23 <0.001
MPA 0.76 ± 0.09 4.57 ± 0.57 0.77 ± 0.15 3.60 ± 0.62 0.79 <0.001
AAo 0.68 ± 0.14 4.66 ± 0.55 0.73 ± 0.18 3.77 ± 0.57 0.31 <0.001
VS 0.44 ± 0.13 4.21 ± 0.92 0.43 ± 0.13 3.42 ± 0.79 0.92 <0.001
Scan time (min) 9.57 ± 5.47 3.62 ± 1.86 <0.001

Values are mean ± standard deviation. Sharpness measure: 0 (blurred) to infinity (sharp). Image quality score: 1 (poor) to 5 (excellent). AAo, ascending aorta; LLPV, lower pulmonary vein; MPA, main pulmonary artery; VS, ventricular septum.