Table 6. Agreement between specimens’ results obtained in the clinical laboratory and results obtained in this study in the same assay.

Assay Total results compared (n) Discordant resultsa (n) Equivocal resultsb (n)
Novatec Novalisa ELISA 2 0 1
DiaSorin LIAISON CLIA 73 0 4
Euroimmun ELISA 49 0 4
Immunetics C6 ELISA 250 4 2
Viramed ViraStripe IB 150 2 14
Euroimmun Euroline IB 101 6 0
Trinity Biotech IB 11 2 2
Seramun SeraSpot 88 0 0
TOTAL 724 14 26

a Discordant: those results negative by the clinical laboratory but positive in the study, or vice versa.

b Equivocal: those results equivocal in the clinical laboratory and positive or negative in this study or vice versa (positive or negative in the clinical laboratory and equivocal in this study).