Skip to main content
. Author manuscript; available in PMC: 2020 May 1.
Published in final edited form as: ASAIO J. 2019 May-Jun;65(4):395–400. doi: 10.1097/MAT.0000000000000918

Table 1.

Summary of Acute Trials

Trial
No.
Cannula Type Implant
Procedure
Device Flow
Rate
[L/min]
Duration
[hr]
Comment
1 18 Fr Straight (ven)1 Left Thoracotomy P-PAL v1 1.0–2.5 4.5 -
16 Fr Straight (art)2

2 22 Fr Right Angle (ven)3 Left Thoracotomy P-PAL v1 1.0–1.4 4 Flow Limitation
16 Fr Straight (art)2 (ven cannula occlusion)

3 22 Fr Right Angle (ven)3 Right Thoracotomy P-PAL v1 1.0–2.5 6 -
16 Fr Straight (art)2

4 22 Fr Right Angle (ven)3 Right Thoracotomy P-PAL v1 1.0–2.4 6 Device Failure
16 Fr Straight (art)2 (↑ bundle resistance)

5 22 Fr Right Angle Metal Tip (ven)4 Right Thoracotomy P-PAL v2 1.0–2.5 6 -
18 Fr Straight (art)5

6 22 Fr Right Angle Metal Tip (ven)4 Right Thoracotomy P-PAL v2 1.0–2.5 6 -
18 Fr Straight (art)5
1

Medtroni DLP Single Stage Venous Cannula (#66118)

2

Medtronic DLP One-Piece Arterial Cannula (#77016)

3

Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula (VCS02290)

4

Medtronic DLP Single Stage Venous Cannula (#69322)

5

Medtronic Elongated One-Piece Arterial Cannula (EOPA) (#77518)