Table 2.
LV leads connected to the RV port | PM/CRTP (n = 22) | ICD/CRTD (n = 15) |
---|---|---|
LV threshold (V@0.4 ms) | ||
Baseline | 0.75 [0.4‐0.8] | 1 [0.7‐1.5] |
FU | 1.15 [0.8‐1.7] | 1.38 [1‐1.5] |
LV signal amplitude (mV) | ||
Baseline | 14.6 [10.7‐19.2] | 11.3 [7.1‐7.8] |
FU | 10 [9.2‐15.7] | 15.25 [9.6‐18.1] |
LV impedance (ohm) | ||
Baseline | 815 [613‐966] | 627 [504‐815] |
FU | 580 [468‐741] * | 665 [585‐798] |
RV leads connected to the LV port (n = 13) | Not applicable | CRTD |
---|---|---|
RV impedance (ohm)
Tip‐coil | ||
At lead switch | 380 [350‐465] | |
FU | 397 [365‐446] | |
SVC coil impendance (ohm) | ||
At lead switch | 59 [55‐62] | |
FU | 60 [55‐61] | |
RV coil impedance (ohm) | ||
At lead switch | 46 [45‐47] | |
FU | 46 [43‐48] | |
RV threshold (V@0.4 ms)
Tip‐coil | ||
At lead switch | 1 [1‐1.3] | |
FU | 1 [1‐1.3] |
Abbreviations: FU, follow‐up; LV, left ventricular; RV, right ventricular; SVC, superior vena cava
P < 0.01.