Performance of LV leads connected to the RV port of CIEDS, and of RV leads connected to the LV port of CRTDs

LV leads connected to the RV port PM/CRTP (n = 22) ICD/CRTD (n = 15)
LV threshold (V@0.4 ms)
Baseline 0.75 [0.4‐0.8] 1 [0.7‐1.5]
FU 1.15 [0.8‐1.7] 1.38 [1‐1.5]
LV signal amplitude (mV)
Baseline 14.6 [10.7‐19.2] 11.3 [7.1‐7.8]
FU 10 [9.2‐15.7] 15.25 [9.6‐18.1]
LV impedance (ohm)
Baseline 815 [613‐966] 627 [504‐815]
FU 580 [468‐741] * 665 [585‐798]
RV leads connected to the LV port (n = 13) Not applicable CRTD
RV impedance (ohm)
Tip‐coil
At lead switch 380 [350‐465]
FU 397 [365‐446]
SVC coil impendance (ohm)
At lead switch 59 [55‐62]
FU 60 [55‐61]
RV coil impedance (ohm)
At lead switch 46 [45‐47]
FU 46 [43‐48]
RV threshold (V@0.4 ms)
Tip‐coil
At lead switch 1 [1‐1.3]
FU 1 [1‐1.3]

Abbreviations: FU, follow‐up; LV, left ventricular; RV, right ventricular; SVC, superior vena cava

*