Skip to main content
. 2019 Jun 7;1594:160–172. doi: 10.1016/j.chroma.2019.02.001

Table 2.

Benefits and constraints of TD-GC×GC-FID/qMS analytical instrumentation.

Instrumentation Benefits Constraints Considerations
Sorbent tubes and thermal desorption autosampler High analytical capacity
Low storage requirements
High storage stability under adequate conditions (brass caps, dry purged, cooled)
Low background artefacts
Maximum carrier pressure 60 psi
Minimum trap flow 2 mL/min
Retention of water on sorbent
Sorbent selectivity and volatility range
Max pressure limits 1D and 2D flow rate combinations
Reduced sensitivity due to necessary low 1D rates creating increased split
Incorporate dry purge step
Flow modulator Inexpensive, low consumables cost
Easy to use, few connections
Suitable for volatile analysis
Requires high 2D flow rate
Broader modulated peak widths
Compatibility of detector(s) with high column flow rates
Acquisition rate of detector(s) to capture modulated peaks
Fast scanning qMS Cost effective
Fast scanning (20,000 amu/s)
Suitable for targeted analysis after biomarker identification
Compatible with < 1.5 mL/min flow rates
Max acquisition rate <40 Hz
Mass spectral skewing
Need to split 2D flow to avoid over-pressurising detector
Setup to either accommodate quantitative or qualitative analysis
Sensitivity changes with acquisition rate and split column flow
FID Cost effective
Compatible with flow modulator
High acquisition rate
Broad dynamic range
No mass spectral assignment Setup to either accommodate quantitative or qualitative analysis
Need supporting mass spectrometer for identification of unknowns
Splitter plate Enables multiple detectors
Easy to use
Offers control of split ratio between detectors
Provides constant flow to one detector
Requires a constant positive pressure from auxiliary EPC
Change in sensitivity due to split
Sensitivity of the secondary detector(s) without constant flow control
Alignment of results requires balanced restrictor lengths