Table 2.
Instrumentation | Benefits | Constraints | Considerations |
---|---|---|---|
Sorbent tubes and thermal desorption autosampler | High analytical capacity Low storage requirements High storage stability under adequate conditions (brass caps, dry purged, cooled) Low background artefacts |
Maximum carrier pressure 60 psi Minimum trap flow 2 mL/min Retention of water on sorbent Sorbent selectivity and volatility range |
Max pressure limits 1D and 2D flow rate combinations Reduced sensitivity due to necessary low 1D rates creating increased split Incorporate dry purge step |
Flow modulator | Inexpensive, low consumables cost Easy to use, few connections Suitable for volatile analysis |
Requires high 2D flow rate Broader modulated peak widths |
Compatibility of detector(s) with high column flow rates Acquisition rate of detector(s) to capture modulated peaks |
Fast scanning qMS | Cost effective Fast scanning (20,000 amu/s) Suitable for targeted analysis after biomarker identification |
Compatible with < 1.5 mL/min flow rates Max acquisition rate <40 Hz Mass spectral skewing |
Need to split 2D flow to avoid over-pressurising detector Setup to either accommodate quantitative or qualitative analysis Sensitivity changes with acquisition rate and split column flow |
FID | Cost effective Compatible with flow modulator High acquisition rate Broad dynamic range |
No mass spectral assignment | Setup to either accommodate quantitative or qualitative analysis Need supporting mass spectrometer for identification of unknowns |
Splitter plate | Enables multiple detectors Easy to use Offers control of split ratio between detectors |
Provides constant flow to one detector Requires a constant positive pressure from auxiliary EPC |
Change in sensitivity due to split Sensitivity of the secondary detector(s) without constant flow control Alignment of results requires balanced restrictor lengths |