Table 2.

Cross-platform testing of the HRM assay.

Virus Genotype Real time PCR machines with Tm values
CFX 96 (Bio-Rad) LC 480 II (Roche) QS 6 (Life tech) RG-Q (Qiagen)
SPPV vaccine 75.8–76.3 76.59–76.88 76.45–76.74 77.2–77.54
SPPV field 78.9–79.4 79.87–80.26 79.74–80.22 80.46–80.9
GTPV 80.4–80.4 81.39–81.52 81.29–81.38 82–82.2
LSDV 80.9–80.9 82.07–82.16 81.96–82.06 82.6–82.66

Four different real-time instruments: the LigthCycler 480 (LC 480 II), QuantStudio 6 (QS 6), RotorGene Q (RG-Q) and CFX 96 were used for assay evaluation. The respective amplicon melting temperature values are indicated for SPPV vaccine and other CaPVs.