Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2018 Sep 19;81(2):1066–1079. doi: 10.1002/mrm.27472

Non‐contrast enhanced simultaneous 3D whole‐heart bright‐blood pulmonary veins visualization and black‐blood quantification of atrial wall thickness

Giulia Ginami 1,, Karina Lòpez 1, Rahul K Mukherjee 1, Radhouene Neji 1,2, Camila Munoz 1, Sébastien Roujol 1, Peter Mountney 1,3, Reza Razavi 1, René M Botnar 1,4, Claudia Prieto 1,4
PMCID: PMC6492092  PMID: 30230609

Abstract

Purpose

Pre‐interventional assessment of atrial wall thickness (AWT) and of subject‐specific variations in the anatomy of the pulmonary veins may affect the success rate of RF ablation procedures for the treatment of atrial fibrillation (AF). This study introduces a novel non‐contrast enhanced 3D whole‐heart sequence providing simultaneous information on the cardiac anatomy—including both the arterial and the venous system—(bright‐blood volume) and AWT (black‐blood volume).

Methods

The proposed MT‐prepared bright‐blood and black‐blood phase sensitive inversion recovery (PSIR) BOOST framework acquires 2 differently weighted bright‐blood volumes in an interleaved fashion. The 2 data sets are then combined in a PSIR‐like reconstruction to obtain a complementary black‐blood volume for atrial wall visualization. Image‐based navigation and non‐rigid respiratory motion correction are exploited for 100% scan efficiency and predictable acquisition time. The proposed approach was evaluated in 11 healthy subjects and 4 patients with AF scheduled for RF ablation.

Results

Improved depiction of the cardiac venous system was obtained in comparison to a T2‐prepared BOOST implementation, and quantified AWT was shown to be in good agreement with previously reported measurements obtained in healthy subjects (right atrium AWT: 2.54 ± 0.87 mm, left atrium AWT: 2.51 ± 0.61 mm). Feasibility for MT‐prepared BOOST acquisitions in patients with AF was demonstrated.

Conclusion

The proposed motion‐corrected MT‐prepared BOOST sequence provides simultaneous non‐contrast pulmonary vein depiction as well as black‐blood visualization of atrial walls. The proposed sequence has a large spectrum of potential clinical applications and further validation in patients is warranted.

Keywords: atrial walls, black‐blood, bright‐blood, pulmonary veins

1. INTRODUCTION

The incidence of atrial fibrillation (AF) is rapidly growing in industrialized countries. AF can be treated with catheter RF ablation procedures aiming at pulmonary vein (PV) isolation. During such interventions, RF ablations are performed around the PV ostia and in regions of atrial tissue characterized by abnormal electrical activity.1 Subject‐specific PV anatomy, PV ostia size, as well as number of PVs are conventionally assessed before the ablation procedure using CT or contrast‐enhanced magnetic resonance angiography (CE‐MRA).2, 3, 4, 5, 6 From these images, static 3D road maps of the PVs and of the atria are generated and used for the guidance of the ablation procedure. Despite the higher spatial resolution provided by CT in comparison to CE‐MRA, such imaging technique typically involves the exposure of the patient to ionizing radiation and iodinated contrast, and is therefore not ideal for repeated imaging of the PVs (e.g., for the assessment of post‐interventional complications or to assess gaps in lesion sets before re‐do ablation). CE‐MRA acquisitions are typically performed during the first‐pass of a contrast agent; patients are asked to hold their breath starting from the arrival of the contrast agent at the level of the PVs and until it reaches the right ventricle. Incorrect synchronization of the patient’s breath‐hold with the CE‐MRA acquisition may result in poor PVs contrast, therefore, and to reduce image blurring because of cardiac motion, free‐breathing ECG‐triggered inversion recovery (IR)‐prepared acquisitions can be performed as an alternative.7 However, the need to determine the subject‐specific optimal inversion time (TI) before acquisition makes this technique intrinsically sensitive to imaging parameters, whereas contrast agent washout can have a detrimental effect on PVs delineation. Non‐contrast enhanced modalities for imaging the PVs have been introduced and rely on the use of a slab‐selective inversion pulse for the enhancement of the PV signal.8, 9 Such techniques, however, are sensitive to both TI and slab thickness, and sophisticated acquisition planning is therefore required. Non‐contrast enhanced imaging of the cardiac venous system—and particularly of the coronary veins and coronary sinus (CS)—has been alternatively achieved using magnetization transfer (MT) contrast (MTC) to ensure high signal from both arteries and veins, while providing adequate contrast between myocardium and blood.10, 11, 12 Free‐breathing protocols for the visualization of the PVs and of the coronary venous system have been introduced in combination with diaphragmatic navigator gating (dNAV) to compensate for respiratory motion, therefore entailing prolonged and unpredictable examination times as only a fraction of the acquired data is accepted for image reconstruction. Furthermore, the use of dNAV provides only an indirect estimation of the respiratory motion of the heart, as it tracks the superior‐inferior (SI) respiratory displacement of the dome of the right hemi diaphragm. In addition, the intersection between the pencil bin navigator and the PVs can cause inflow artefacts and lead to reduced sharpness of such vessels.7, 13, 14 Therefore, the development of a non‐contrast enhanced sequence for imaging the cardiac venous system that operates with predictable examination time and without entailing sophisticated acquisition planning would be highly desirable.

Nevertheless, and despite pre‐interventional assessment of the PV anatomy being routinely performed, many patients treated with catheter ablation require repeated procedures over time, as the success rate of these interventions remains suboptimal.15 Atrial wall thickness (AWT) plays a major role in atrial arrhythmogenesis and its pre‐interventional assessment may affect the success rate of RF ablation procedures.16 In fact, electrical reconnection of the ablated tissues may occur when insufficient thermal energy is delivered, therefore preventing the creation of truly transmural and electrically isolating lesions.15 Conversely, excessive energy delivery increases the risk of atrial perforation, tamponade, and atrio‐oesophageal fistula formation.15, 17, 18 Furthermore, it has been recently demonstrated that AF tends to originate from regions characterized by highly varying AWT.19 Therefore, AWT quantification may help in identifying the optimal energy level to be delivered during ablation as well as the most favorable ablation targets. CT has been used for in vivo quantification of AWT20, 21, 22, 23; however, reduced soft tissue contrast obtained with this imaging modality can challenge the detection of extensive portions of the atrial walls. More recently, the use of MRI has been proposed for the quantification of AWT24; with this approach, black‐blood MRI of the atrial walls is obtained exploiting a phase‐sensitive inversion recovery (PSIR) sequence25 alternating the acquisition of a long inversion time (TI) black‐blood image and of a low flip‐angle reference image used for phase computation. Such acquisition scheme, however, is integrated with dNAV for respiratory motion compensation, therefore entailing all the previously described drawbacks. Furthermore, the acquisition of a low flip‐angle reference image doubles the scan time without providing any additional diagnostic information. In addition, black‐blood imaging alone is not ideal for the visualization of the overall cardiac anatomy, including the PVs.

This study aims at introducing a novel MRI sequence suitable for non‐contrast enhanced interventional planning of atrial ablation procedures providing (1) a 3D whole‐heart bright‐blood volume for comprehensive visualization of the cardiac anatomy, and (2) a complementary, co‐registered, 3D whole‐heart black‐blood volume from which AWT measurements can be derived. The proposed sequence is integrated with image‐based navigation26 and non‐rigid respiratory motion correction,27 leading to predictable acquisition time, nearly 100% scan efficiency, and allowing for the estimation of respiratory motion parameters from the heart directly, thereby avoiding the use of motion models. Such approach was validated in healthy subject and patients who were scheduled for an RF ablation procedure to assess its clinical feasibility.

2. METHODS

2.1. Framework implementation

A previously published28 ECG‐triggered 3D whole‐heart bright‐blood and black‐blood PSIR bSSFP (BOOST) sequence was adapted to perform data acquisition as illustrated in Figure 1. The sequence exploits a Cartesian trajectory with spiral profile order,29 where consecutive spiral interleaves are rotated by the golden angle. The implementation of the BOOST sequence described in Ginami et al.28— originally proposed for the simultaneous pre‐contrast visualization of the coronary artery lumen (bright‐blood) and of coronary thrombus and intraplaque hemorrhage (black‐blood) — alternates the acquisition of 2 differently weighted bright‐blood volumes (T2Prep‐IR BOOST and T2Prep BOOST) that are subsequently combined in a PSIR‐like reconstruction to obtain a third, complementary, and fully co‐registered black‐blood volume. Because of the detrimental effect of the T2Prep pre‐pulse on coronary vein depiction,10 this sequence implementation might not be directly applicable for the visualization of the cardiac venous system. Therefore, we introduce a novel BOOST configuration exploiting MT‐preparation instead of T2Prep with the goal of enabling a sharp delineation of both the cardiac arterial and venous systems. Furthermore, non‐rigid motion correction is incorporated in the proposed MT‐prepared BOOST, in contrast to the previously published T2‐prepared BOOST sequence28 that considers only translational motion during the process of image reconstruction. The BOOST implementation proposed in this study alternates the acquisition of an IR pulse preceded by MT preparation (bright‐blood MTC‐IR BOOST, Figure 1A), whereas MT preparation solely (bright‐blood MTC BOOST, Figure 1B) is applied at even heartbeats. In odd heartbeats, a short TI IR approach30 is used to suppress the signal from epicardial fat (Figure 1A), whereas frequency‐selective pre‐saturation31 is used in even heartbeats (Figure 1B). In each heartbeat, a low resolution 2D image‐based navigator (iNAV) is acquired by spatially encoding the ramp‐up pulses of the bSSFP sequence, therefore allowing for the estimation of respiratory motion along the SI and right‐left (RL) directions. Respiratory motion estimation and compensation is performed in a beat‐to‐beat (2D translational) and bin‐to‐bin (3D non‐rigid) fashion for the 2 bright‐blood data sets (MTC‐IR BOOST and MTC BOOST) independently and before PSIR computation. A rectangular region of interest (ROI) corresponding to the iNAV is manually selected covering the whole heart along the RL direction and the base and the mid part of the heart along the SI direction. The beat‐to‐beat translational respiratory motion of the heart is then estimated along the SI and RL directions using a template‐matching algorithm exploiting cross‐correlation32; respiratory motion correction is performed toward the end‐expiratory level.33 The estimated SI displacement is then used to assign each data segment to a specific bin, corresponding to its position within the respiratory cycle; bins are generated to be equally populated. Respiratory outliers (falling outside the interval represented by mean ± 2 times the SD of the estimated SI respiratory signal) are rejected from the process of 3D non‐rigid motion corrected reconstruction. Data belonging to each respiratory bin are translationally corrected to the center of the bin itself using the previously estimated SI and RL displacements by applying a linear phase shift in k‐space.26 Subsequently, respiratory‐resolved bins are reconstructed using soft‐gated iterative SENSE (It‐SENSE)34 with exponential decay weighting.34, 35, 36 Once each individual respiratory bin is reconstructed, the most end‐expiratory bin is used as reference to estimate bin‐to‐bin 3D non‐rigid deformation fields via image registration as described in Modat et al.37 The estimated 3D non‐rigid motion fields are then incorporated in a generalized matrix description reconstruction pipeline where the motion‐corrected 3D volume is reconstructed using a linear conjugate gradient method.27, 29, 38 The non‐rigid end‐expiratory motion corrected MTC‐IR BOOST and MTC BOOST data sets are then combined in a PSIR‐like reconstruction25 to obtain a complementary black‐blood data set (PSIR BOOST, Figure 1C); in this process, the MTC BOOST volume is used as the reference image for phase computation.

Figure 1.

Figure 1

Proposed framework for simultaneous 3D whole‐heart bright‐blood depiction of the PVs and heart anatomy and black‐blood visualization of atrial walls. Two magnetization prepared bright‐blood volumes are acquired in odd and even heartbeats. Specifically, magnetization transfer in combination with an inversion pulse is used in odd heartbeats (MTC‐IR BOOST A), whereas magnetization transfer solely is exploited in even heartbeats (MTC BOOST B), The MTC‐IR acquisition is designed for comprehensive visualization of the heart anatomy. Although a short TI is exploited for fat saturation in odd heartbeats, spectral presaturation (Fat Sat) is used in even heartbeats. Data acquisition is performed using a 3D Cartesian trajectory with spiral profile order and segmented over multiple heartbeats (green, red, blue). A low‐resolution 2D iNAV is acquired in each heartbeat by spatially encoding the ramp‐up pulses of the bSSFP sequences. The bright‐blood MTC‐IR BOOST and MTC BOOST volumes are non‐rigidly motion corrected at the end‐expiratory level and, subsequently, combined in a PSIR‐like reconstruction to generate a complementary black‐blood volume for atrial wall visualization (PSIR BOOST, C). PVs, pulmonary veins; MTC, magnetization transfer contrast; IR, inversion recovery pulse; PSIR, phase sensitive inversion recovery

2.2. Data acquisition—healthy subjects

The proposed prototype framework was tested in 11 healthy subjects (29.6 ± 3.1 y, 5 males) on a 1.5T MRI system (Magnetom Aera, Siemens Healthcare, Erlangen, Germany). The study was approved by the Institutional Review Board and written informed consent was obtained for each subject. Data were acquired using an 18‐channel chest coil and a 32‐channel spine coil. Relevant imaging parameters for the proposed MT‐prepared BOOST sequence included: coronal orientation, isotropic spatial resolution 1.4 × 1.4 × 1.4 mm3, subject specific FOV = 320 × 320 × 90–120 mm3, TE = 1.4 ms, TR = 3.5 ms, pixel bandwidth = 1395 Hz/pixel, flip‐angle for both odd and even heartbeats = 90°. For both the MTC‐IR BOOST and the MTC BOOST images, MT preparation consisted of 15 off‐resonance Gaussian pulses optimized to prevent on‐resonance saturation with the following parameters: bandwidth‐time‐product (BWTP) = 1.92, flip angle = 800°, pulse duration = 20.48 ms, off‐resonance frequency offset = 3000 Hz, pause between pulses = 1.5 ms. Fourteen bSSFP ramp‐up pulses were played out, leading to a spatial resolution of 1.4 × 22.8 mm2 (interpolated to 1.4 × 1.4 mm2 before motion estimation) for each iNAV. The TI was set to 140 ms for odd heartbeats. Additionally, and for comparison purposes, the previously published T2‐prepared BOOST sequence28 was acquired with matching imaging parameters. For both odd (T2Prep‐IR BOOST) and even (T2Prep BOOST) heartbeats, T2Prep duration was set equal to 40 ms, whereas the TI was set to 110 ms for odd heartbeats. Before data acquisition, a high‐resolution Cine was acquired in transversal orientation to determine the subject specific mid‐diastolic resting period (typically ranging between 95–120 ms, corresponding to 28–35 k‐space lines acquired per heartbeat). For 1 individual volunteer (subject 2), and for illustration purposes, a 3D whole‐heart PSIR sequence25 was acquired using an imaging protocol similar to that previously introduced for MRI‐derived measurements of AWT.24 Briefly, an IR pulse was applied in odd heartbeats, where data acquisition was performed with a 90° flip angle. The TI was set equal to 360 ms to null the signal from blood, as determined using a 2D TI scout scan that was acquired in coronal view. The acquisition of the reference image in even heartbeats was performed without preparatory pulses and at a low flip‐angle, set to 8°. To allow for respiratory motion estimation and correction, iNAVs were acquired both in odd and even heartbeats, whereas in the original publication, dNAV was used.24 Spatial resolution and other imaging parameters were set to match those of the BOOST acquisitions.

2.3. Data acquisition—patients

Data acquisition was additionally performed in 4 patients (58.3 ± 6.4 years old, 3 males) to test the feasibility of the MT‐prepared BOOST sequence in a clinical setting. Patients were presenting with persistent AF and were scheduled for a RF ablation procedure. The proposed MT‐prepared BOOST sequence was acquired pre‐contrast administration with imaging parameters matching those of the healthy subject acquisitions except for the spatial resolution that was set to 1.4 × 1.4 × 2.8 mm3 and then interpolated to 1.4 × 1.4 × 1.4 mm3 during the process of image reconstruction. After the acquisition of the MT‐prepared sequence, contrast was injected (0.2 mmol/kg, Gadovist, Bayern, Berlin, Germany). The clinical protocol included the acquisition of a post‐contrast (starting 90 s after injection) IR‐prepared ECG‐triggered gradient‐echo (GRE) sequence, for the assessment of the PV anatomy, with the following imaging parameters: spatial resolution 1.3 × 1.3 × 2 mm3, subject‐specific FOV covering the left atrium along the SI direction = 332 × 332 × 60–90 mm3, transversal orientation, TE = 1.8 ms, TR = 8 ms, pixel bandwidth = 370 Hz/pixel, flip angle 20°, TI = 200 ms (nulling the myocardial signal). The bright‐blood clinical sequence was integrated with dNAV for respiratory motion suppression. An acceptance window of ±3.5 mm was placed in end‐expiration, and a GRAPPA39 acceleration factor of 2 was used with 24 reference lines. All patients were in AF at the time of image acquisition.

2.4. Image reconstruction

Raw data were retrieved from the scanner and reconstructed in MATLAB (The MathWorks, Natick, MA). For the proposed MT‐prepared BOOST sequence, MTC‐IR BOOST and MTC BOOST data were independently reconstructed in end‐expiration using the previously described non‐rigid motion correction framework.27 A number of equally populated bins ranging from 4 to 6 were reconstructed for each individual data set, considering a maximum bin size of 3.5 mm. A fixed and empirically optimized number of 5 iterations was used for the It‐SENSE reconstructions (reconstruction of the respiratory bins and of the final, non‐rigid motion corrected images), involving ~95–100% of the acquired data after outlier rejection. Additionally, volunteer data were also reconstructed with beat‐to‐beat 2D translational motion correction only and without motion correction for comparison purposes. The non‐rigid motion corrected MTC‐IR BOOST and MTC‐BOOST data sets were combined in a PSIR‐like reconstruction as described in Kellman et al.,25 generating a third, complementary, PSIR BOOST data set for atrial wall visualization. Volunteer data acquired with the previously published T2‐prepared BOOST sequence28 and PSIR reconstruction resembling that described in Varela et al.24 were also reconstructed using the non‐rigid motion correction pipeline. For the patient acquisitions, the bright‐blood clinical sequence was reconstructed using the scanner software (Syngo MR E11A, Siemens Healthcare).

2.5. Data analysis—motion correction

Image quality measurements were performed for the healthy subjects and on the bright‐blood MTC‐IR BOOST data sets reconstructed with non‐rigid motion correction, translational motion correction only, and without motion correction to assess the performance of non‐rigid respiratory motion correction in combination with the BOOST framework. Specifically, visible vessel length (VL) and percentage vessel sharpness (%VS) were computed using the software described in Etienne et al.40 along the right inferior (rPV) and the left inferior (lPV) pulmonary veins. Furthermore, %VS and VL were computed for the following coronary vessels: anterior interventricular coronary vein (AIV), posterior interventricular coronary vein (PIV), right coronary artery (RCA), and left anterior descending coronary artery (LAD). All the %VS and VL measurements were computed along the first visible 4 cm for all the vessels. Quantitative parameters were compared with a paired 2‐tailed Student’s t‐test; Bonferroni correction for multiple comparisons was applied, resulting in a P‐value of 0.016 as the threshold for statistical significance.

2.6. Data analysis—vein visualization

The bright‐blood MTC‐IR BOOST data sets and the bright‐blood T2Prep‐IR BOOST data sets acquired in healthy subjects, were compared to assess their ability to depict the cardiac venous and arterial systems. SNR of both venous (SNRven) and arterial (SNRart) blood were computed for both data sets, together with contrast‐to‐noise ratio (CNR) of both venous (CNRven) and arterial (CNRart) blood with respect to the myocardium. For SNR and CNR measurements, ROIs were selected at the level of the CS (venous blood), left ventricle (arterial blood), septum (myocardium), and lungs (background). Furthermore, VL and %VS of rPV, lPV, AIV, PIV, and RCA were computed for the non‐rigid motion corrected T2Prep‐IR BOOST data sets and compared to the corresponding values quantified for the non‐rigid motion corrected MTC‐IR BOOST data sets. A paired 2‐tailed Student’s t‐test was performed to assess statistical significance for all quantitative endpoints, assuming P < 0.05 as statistically significant.

2.7. Data analysis—AWT quantification

Measurements of AWT were performed on the black‐blood PSIR BOOST data sets obtained using the proposed MT‐prepared BOOST sequence in healthy subjects. For each individual black‐blood PSIR BOOST data set, a 4‐chamber view—perpendicular to both the right and the left atrium—was generated using the Horos software (V1.1.7). Resulting images were then loaded in the Soap‐Bubble software40 to quantify AWT for both right and left atria. Furthermore, to demonstrate feasibility of atrial wall segmentation with the proposed framework, individual slices oriented along the coronal, transversal, and sagittal views were processed using a semi‐automatic machine‐learning based Trainable Weka Segmentation (TWS) algorithm41 implemented in the Fiji software (V3.2.20).42 For each processed slice, 3 different ROIs were manually selected in blood, background, and myocardium and/or wall regions to train the classifier. Segmentation results were overlaid onto the original black‐blood PSIR BOOST image for visualization purposes.

3. RESULTS

Data acquisition and reconstruction were carried out successfully for all subjects. All free‐breathing sequences were acquired in healthy subjects with predictable nominal acquisition time of ~16–18 min. Nominal examination time for BOOST acquisitions in patients was ~7–8 min (to obtain both the whole‐heart bright‐ and the black‐blood volume). The bright‐blood clinical sequence (providing volumetric coverage of the left atrium) was acquired in ~5 min, with an average acquisition efficiency of ~50%.

3.1. Motion correction

iNAVs preceding both the bright‐blood MTC‐IR BOOST and MTC BOOST volumes allowed to extract respiratory motion information from the heart directly (Figure 2A,B). This enabled non‐rigid motion corrected reconstruction of bright‐blood data sets for the visualization of the heart anatomy (odd heartbeats) as well as reconstruction of a high‐quality black‐blood PSIR BOOST data set for AWT quantification (Figure 2E). Conversely, the use of image‐based navigation in combination with the conventional PSIR sequence resembling that described in Varela et al.24 was more challenging. In fact, cardiac visualization was poor in iNAVs acquired in odd heartbeats, were the TI was set to null the blood signal (Figure 2C). Therefore, the iNAVs tracked the respiratory motion of the high‐contrast heart–liver interface, such that direct estimation of respiration induced cardiac motion cannot be attained. Additionally, iNAVs acquired in even heartbeats exhibit reduced signal because of the low flip‐angle used for the imaging readout (Figure 2D). Furthermore, the low signal of the heart in both odd and even heartbeats challenges the estimation of the non‐rigid motion fields required for the second step of the motion correction pipeline, and this results in a low‐quality black‐blood data set with compromised heart sharpness in the final PSIR reconstruction (Figure 2F). For the proposed MT‐prepared BOOST framework, bright‐blood MTC‐IR BOOST data sets corrected with 2D translational motion exhibit higher overall image quality when compared to the motion corrupted counterparts. However, as the iNAV was typically placed at the inferior part of the heart to estimate the global translational motion, structures that were not included in the tracking template exhibit reduced sharpness in some cases (Figure 3A,B,D, and E). Conversely, the use of 3D non‐rigid motion correction uniformly improved image quality and delineation of the PVs, the coronary vessels, and the myocardium, as statistically quantified (Table 1) and as can be visually appreciated for 2 individual subjects in Figure 3C,F. All quantified endpoints for uncorrected, 2D translational motion‐corrected, and 3D non‐rigid motion‐corrected bright‐blood MTC‐IR BOOST data sets are summarized in Table 1.

Figure 2.

Figure 2

Image‐navigated respiratory motion tracking with the proposed MT‐prepared BOOST configuration and with a more conventional approach for black‐blood PSIR. The BOOST framework acquires 2 differently weighted bright‐blood data sets (MTC‐IR BOOST and MTC BOOST), providing iNAVs where the heart is depicted with high signal and contrast (A and B). As such, the respiratory displacement can be extracted using a template positioned at the level of the heart itself (red rectangles). This leads to a sharp depiction of the cardiac anatomy after non‐rigid motion correction. Similarly, this results in good anatomy depiction after PSIR computation (E). In contrast, and with a more conventional black‐blood PSIR sequence, iNAVs exhibit low signal and contrast because of blood signal nulling (C) and to a low flip‐angle acquisition for the reference image (D). Consequently, the respiratory motion appears to be tracked at the level of the high‐contrast interface between the liver and the heart and lungs. Although this leads to a sharp delineation of the liver and abdominal vessels, the low signal in both odd and even heartbeats prevents the estimation of non‐rigid motion fields at the level of the heart; this results in reduced heart sharpness after PSIR computation (F). PSIR, phase sensitive inversion recovery; MTC, magnetization transfer contrast; IR, inversion recovery pulse; iNAV, image‐based navigator; UC, uncorrected

Figure 3.

Figure 3

Motion correction performances using 2D translational and 3D non‐rigid motion correction in 2 representative healthy subjects. Before motion correction (red rectangle, A and D), images are affected by severe motion blurring preventing a sharp visualization of the overall cardiac anatomy. Translational motion correction (blue rectangles, B and E) already improves the depiction of several anatomical details, especially in the area where the iNAV template is typically placed (green arrows). Differently, static structures or structures without the template suffer from introduced blurring (yellow arrows). The use of non‐rigid respiratory motion correction successfully restores overall image sharpness, leading to the visualization of the overall cardiac anatomy with excellent image quality (purple rectangles, C and F)

Table 1.

Pulmonary vein, coronary veins, and arteries depiction for different motion correction strategies

Uncorrected Translational Non‐rigid
rPV VL (cm) 4.09 ± 1.90 4.70 ± 2.22 5.51 ± 2.02a
%VS 42.45 ± 16.96 48.43 ± 17. 25a 56.90 ± 6.24a
lPV VL (cm) 4.88 ± 3.19 4.28 ± 2.01 6.08 ± 3.20b
%VS 35.03 ± 18.78 46.95 ± 17.04 60.16 ± 4.83a, b
AIV VL (cm) 1.70 ± 1.58 2.72 ± 1.44 4.92 ± 2.03a, b
%VS 25.12 ± 24.49 48.11 ± 8.00a 52.17 ± 4.88a
PIV VL (cm) 1.58 ± 1.38 3.61 ± 2.36a 5.11 ± 3.35a
%VS 24.24 ± 20.13 44.19 ± 7.78a 51.49 ± 8.72a, b
LAD VL (cm) 6.83 ± 4.89 8.58 ± 4.21 10.75 ± 5.20a, b
%VS 39.73 ± 15.34 47.93 ± 7.57a 57.38 ± 6.66a, b
RCA VL (cm) 5.30 ± 2.89 6.70 ± 2.68a 6.70 ± 2.89a
%VS 39.12 ± 21.35 54.27 ± 8.31a 58.83 ± 6.62a

Abbreviations: rPV, right inferior pulmonary vein; IPV, left inferior pulmonary vein; AIV, anterior interventricular vein; PIV, posterior interventricular vein; LAD, left anterior descending coronary artery; RCA, right coronary artery; VL, visible vessel length; %VS, percentage vessel sharpness.

Although translational motion corrected data sets showed improved image quality when compared to the uncorrected counterparts, the use of non‐rigid respiratory motion correction further improves coronary vein and artery visualization.

a

Indicates significant improvement with respect to the uncorrected data sets.

b

Indicates significant improvement with respect to translational motion correction only. P < 0.016 is considered the threshold for statistical significance, as per Bonferroni correction for multiple comparisons.

3.2. Vein visualization

The use of MT preparation in the MTC‐IR BOOST data sets enabled the depiction of both venous and arterial blood with high signal and contrast. Conversely, the use of T2Prep in T2Prep‐IR BOOST resulted in a detrimental effect on the visualization of the cardiac venous system. Statistically improved (P < 0.05) SNRven and CNRven were quantified in the MTC‐IR BOOST data sets in comparison to the T2Prep‐IR BOOST counterparts. Furthermore, the use of MT preparation resulted in a statistically higher SNRart. On average, higher CNRart was quantified in T2Prep‐IR BOOST, yet without being significantly different from the CNRart that was quantified in MTC‐IR BOOST. Higher SNRven and CNRven resulted in statistically improved pulmonary and coronary vein delineation in MTC‐IR BOOST; in fact, increased %VS and VL were quantified for rPV, lPV, AIV, and PIV in comparison to T2Prep‐IR BOOST. The 2 bright‐blood volumes, obtained with the MT‐prepared BOOST sequence proposed in this study and with the previously published T2‐prepared BOOST,28 provided comparable coronary artery delineation in terms of %VS and VL for both RCA and LAD. Figure 4 shows arterial and venous systems depiction obtained with MTC‐IR BOOST and with T2Prep‐IR BOOST in 2 representative subjects. All endpoints quantified to assess venous and arterial blood signal and contrast as well as pulmonary and coronary vein and artery visualization with both sequences are summarized in Table 2. Images obtained in patients for both MTC‐BOOST and the clinical bright‐blood sequences are displayed for 3 patients in Figure 5. PVs are shown for both MTC‐BOOST and clinical bright‐blood sequences; MTC‐BOOST images were reformatted to a different imaging plane than the acquired clinical one, because of the different acquisition orientations and non‐isotropic resolutions. Corresponding anatomical structures could be identified by both sequences, as indicated by the arrows in Figure 5. Reformats of the 2 sequences in both coronal and transversal orientation are provided in Supporting Information Figure S1.

Figure 4.

Figure 4

Arterial and venous systems visualization with MTC‐IR BOOST and with T2Prep‐IR BOOST in 2 representative healthy subjects. The proposed MTC‐IR BOOST approach provides comparable signal intensity for both arterial and venous blood (A, B, E, and F). As a result, the pulmonary veins (red arrow), arteries (yellow arrows), CS (green arrows), and both coronary arteries and veins (blue arrows) are depicted with high contrast. In contrast, the use of T2Prep in T2Prep‐IR BOOST has a detrimental effect on the venous blood signal (C, D, G, and H). Although the coronary arteries appear sharply depicted (blue plain arrows), the pulmonary veins (red arrows), the CS (green arrows) and the coronary veins (blue dotted arrows) are visualized with reduced signal and contrast. CS, coronary sinus; LAD, left anterior descending coronary artery; AIV, anterior interventricular vein; MTC, magnetization transfer contrast; IR, inversion recovery pulse; T2Prep, T2‐preparation

Table 2.

Quantified endpoints comparing the MT‐prepared BOOST sequence proposed in this study and its T2‐prepared counterpart in terms of bright‐blood visualization of pulmonary veins, coronary veins, and arteries

MTC IR BOOST T2Prep IR BOOST P‐value
SNR ven 18.78 ± 7.11 9.97 ± 4.66 <0.0001
SNR art 19.92 ± 7.28 17.03 ± 6.79 <0.002
CNR ven 9.36 ± 3.10 5.22 ± 3.61 <0.0001
CNR art 10.50 ± 3.54 12.28 ± 4.80 NS
rPV VL (cm) 5.51 ± 2.02 1.03 ± 1.80 <0.001
%VS 56.90 ± 6.24 15.56 ± 26.77 <0.001
lPV VL (cm) 6.08 ± 3.20 1.36 ± 2.38 <0.001
%VS 60.16 ± 4.83 15.72 ± 27.00 <0.001
AIV VL (cm) 4.92 ± 2.03 1.87 ± 2.61 <0.001
%VS 52.17 ± 4.88 26.64 ± 25.72 <0.01
PIV VL (cm) 5.11 ± 3.35 1.59 ± 2.25 <0.01
%VS 51.49 ± 8.72 22.56 ± 26.09 <0.01
LAD VL (cm) 10.75 ± 5.20 9.20 ± 4.32 NS
%VS 57.38 ± 6.66 60.05 ± 7.06 NS
RCA VL (cm) 6.70 ± 2.89 6.34 ± 2.34 NS
%VS 58.83 ± 6.62 60.58 ± 5.60 NS

Abbreviations: CNR, contrast to noise ratio; rPV, right inferior pulmonary vein; IPV, left inferior pulmonary vein; AIV, anterior interventricular vein; PIV, posterior interventricular vein; LAD, left anterior descending coronary artery; RCA, right coronary artery; VL, visible vessel length; %VS, percentage vessel sharpness.

The 2 sequences provided comparable arterial blood signal and contrast, leading to similar quality in terms of coronary artery visualization. Conversely, poor signal and contrast of venous blood is obtained with T2Prep‐IR BOOST, therefore leading to degraded depiction of both pulmonary and coronary veins. In contrast, the use of MT‐preparation in MTC‐IR BOOST preserves venous blood signal and contrast, leading to significantly improved depiction of the pulmonary and coronary veins with respect to T2Prep‐IR BOOST.

Figure 5.

Figure 5

Images obtained in patients presenting with AF and scheduled for an ablation procedure. The contrast‐enhanced clinical sequence is displayed as a maximum intensity projection image in (A), (D), and (G). Non‐contrast enhanced BOOST images are displayed as reformats in (B), (E), and (H) (bright‐blood) and in (C), (F), and (I) (black‐blood). Corresponding anatomical structures at the level of the PVs can be visualized in both images (colored arrows). PV, pulmonary veins

3.3. AWT quantification

The PSIR BOOST data sets obtained with the framework proposed in this study showed effective blood signal suppression, whereas adequate contrast could be observed at the level of both the blood–myocardium/walls and the myocardium/wall–lung interfaces (Figures 6 and 7). Quantified AWT thickness amounted to 2.54 ± 0.87 mm for the right atrium and to 2.51 ± 0.61 mm for the left atrium, which is in good agreement with MRI‐derived measurements previously obtained in healthy subjects (2.7 ± 0.7 mm and 2.4 ± 0.7 mm for right and left atrium, respectively).24 Results of successful segmentation obtained from the TWS algorithm are displayed in Figure 6 for 3 representative subjects and for coronal, transversal, and sagittal views. Figure 7 shows segmentation results obtained in 1 subject (subject 11) from the PSIR MT‐prepared BOOST and the PSIR T2‐prepared BOOST volumes. The MT‐prepared BOOST sequence leads to a myocardial and atrial wall signal intensity that is between that of blood and surrounding lungs in the bright‐blood MTC‐IR BOOST data set (Figure 7A). For this reason, the myocardium and the walls appear depicted with good contrast with respect to both blood and lungs in the complementary black‐blood PSIR reconstruction (Figure 7B). This facilitates the process of segmentation that successfully discriminates between the different tissues as displayed in Figure 7C. In contrast, a more pronounced suppression of the myocardial signal is obtained in the bright‐blood T2Prep‐IR BOOST volume (Figure 7D), resulting in a very poor contrast between the myocardium and the lungs in the final black‐blood PSIR reconstruction (Figure 7E). This challenges the segmentation process at the level of such interface (Figure 7F).

Figure 6.

Figure 6

Atrial wall segmentation computed in the PSIR BOOST data set obtained with the proposed MT‐prepared BOOST configuration in 3 representative healthy subjects. Segmentation results (red lines) are displayed along the coronal (A, D, and G), transversal (B, E, and H), and sagittal views (C, F, and I). The myocardium, atrial, and ventricular walls are depicted with high contrast in comparison to both blood and surrounding lungs. This allows the segmentation algorithm to successfully discriminate between the 3 different tissues and to successfully track anatomical borders

Figure 7.

Figure 7

Comparison of atrial wall visualization and segmentation with the proposed MT‐prepared BOOST sequence and with the previously introduced T2‐prepared BOOST configuration. In the bright‐blood MTC‐IR volume (A), designed for the visualization of arterial and venous system (the red arrow indicates one of the pulmonary veins), myocardium and walls exhibit a signal intensity that is intermediate between the one of the blood and that of the lungs (purple arrow). Therefore, intermediate signal intensity of myocardium and walls is preserved in the derived PSIR reconstruction (B); this leads to successful segmentation of myocardium and walls (C, red lines and zoomed areas in blue and yellow boxes). Conversely, in the bright blood T2‐Prep volume (D), venous vessels are not visible (purple arrow) and the myocardial signal is pronouncedly suppressed (red arrow). Consequently, the derived black‐blood PSIR image exhibits reduced contrast between the myocardium and/or walls and the surrounding lungs (E). Although the high‐contrast interface between the blood and the myocardium is successfully recognized and tracked, the segmentation algorithm fails in discriminating between the myocardium and/or walls and the surrounding lungs because of poor contrast at the level of that tissue interface (F, red lines and zoomed areas in blue and yellow boxes)

4. DISCUSSION

4.1. Technical novelties

In this study, we introduce a free‐breathing 3D whole heart MT‐prepared BOOST framework that is suitable for non‐contrast enhanced interventional planning of atrial ablation procedures which provides simultaneously (1) a bright‐blood 3D whole‐heart volume allowing for the visualization of the heart anatomy, including the PVs, and (2) a complementary co‐registered black‐blood 3D whole‐heart volume allowing for atrial wall depiction and AWT quantification. In contrast to previously published approaches for non‐contrast enhanced visualization of the cardiac venous system10, 11, 12 and atrial walls,24 the proposed framework is integrated with image‐based navigation and non‐rigid respiratory motion compensation. This allows for data acquisition with nearly 100% scan efficiency and with predictable nominal acquisition time. The BOOST framework, as previously demonstrated,28 and as it was further shown in this study (Figure 2), is particularly suitable for being integrated with image‐based navigation. In fact, BOOST relies on the acquisition of 2 differently weighted bright‐blood volumes from which respiratory motion information can be independently extracted using bright‐blood iNAVs where the heart is depicted with high signal and contrast. The complementary black‐blood volume (PSIR BOOST) is then obtained during post‐processing via PSIR computation. As BOOST can achieve ~100% scan efficiency, our framework not only provides complementary information on both cardiac anatomy (bright‐blood MTC‐IR BOOST) and AWT (black‐blood PSIR BOOST), but in addition data acquisition is twice as fast as a 3D whole‐heart PSIR sequence with 50% dNAV efficiency and the same acceleration factor. This increased scan efficiency of our framework allows for reduced examination times, which may be beneficial in patients who are prone to respiratory irregularities.43 Our previously introduced T2‐prepared BOOST was found to provide degraded depiction of the cardiac venous system and of the PVs; therefore, such a sequence cannot be directly used for the planning of RF ablation procedures. Conversely, the MT‐prepared BOOST sequence enables good vein delineation.28 In this study, a high frequency offset (3000 Hz) was chosen for MT preparation to minimize image artefacts in regions with imperfect B0 and shimming and to minimize on‐resonance saturation.10 To maintain effective myocardial suppression using such high frequency offset, 15 pulse repetitions and a high flip angle of 800° were used, which can result in increased SAR; however, as the pulses that were used for MT‐preparation were relatively long, the SAR that was computed by the scanner software (1.2–1.6 W/kg, depending on the weight and heartrate of the subject) was within the safety limits for each individual acquisition that was performed in this study. Furthermore, the MT‐prepared BOOST sequence proposed in this study is more suitable not only for pulmonary and coronary veins depiction (bright‐blood MTC‐IR BOOST) but also for atrial wall segmentation and AWT quantification (black‐blood PSIR BOOST) when compared to the T2‐prepared counterpart (Figure 7). The original T2‐prepared BOOST sequence28 considers translational respiratory motion solely in the process of image reconstruction; in contrast, this study integrates BOOST with non‐rigid respiratory motion correction. Consistently with previous findings,27, 44 the use of non‐rigid respiratory motion correction significantly improves image quality when compared to translational motion correction only; this aspect may be particularly beneficial for a future wider validation in patients. The BOOST sequence exploits a bSSFP readout for imaging data acquisition. Although previous studies suggest that the use of a spoiled gradient echo (GRE) readout might be preferable in combination with MTC,10 bSSFP typically provides better delineation of vessels and anatomic structures; this aspect may facilitate the visualization of thin structures like the atrial walls. However, and although good image quality was obtained with the proposed framework, a one‐to‐one comparison between GRE and bSSFP for the proposed BOOST sequence was not performed in this study (mainly to limit overall scan time) and will be investigated in future work.

4.2. Study limitations

In this study, a direct comparison between the proposed framework and other MRI sequences in terms of quantified AWT was not performed. The acquisition of the 2 different BOOST sequences (the one proposed in this study and that introduced in Ginami et al.)28 required ~45 min of scan time, including acquisition planning. Therefore, the acquisition of additional sequences for AWT quantification was impractical. A comparison between the MT‐prepared BOOST and the previously described MRI protocol for AWT measurements24 was performed in 1 subject only, for illustration purposes; the goal of this experiment was to show that the combination of more conventional black‐blood PSIR sequences with image‐based navigation as well as with non‐rigid respiratory motion correction appears to not be straightforward, because of the low signal from the heart in the acquired high resolution images and in the corresponding iNAVs. In this study, feasibility for MT‐prepared BOOST acquisitions in patients was shown; both the non‐contrast enhanced MTC‐IR BOOST data set and the contrast enhanced clinical sequence could provide the depiction of the pulmonary vein with high‐contrast, and corresponding anatomical structures could be visualized in both data sets. However, as the sequences were acquired with different orientations and resolutions, a rigorous one‐to‐one quantitative comparison could not be performed and will be investigated in future studies.

4.3. Future perspectives

The acquisition time of the fully sampled MT‐prepared BOOST was ~16–18 min. Acquisition time was reduced to 7–8 min in patient acquisitions by increasing the slice thickness that may, however, challenge the depiction of the smallest structures. Therefore, the proposed framework will be combined with acceleration techniques, exploiting an implementation similar to that described in Correia et al.,36 where non‐rigid respiratory motion correction of isotropic 3D whole‐heart data sets is integrated with image reconstruction from undersampled data. This will allow the acquisition of high‐resolution data sets within a clinically feasible scan duration.15, 45 In this study, the feasibility of the MT‐prepared BOOST sequence in a clinical setting was showed in 4 patients. A broader clinical validation of the proposed framework—once further technical developments including accelerated image acquisition will have been implemented—is foreseen; this MT‐prepared BOOST implementation will be tested in a larger population of patients undergoing atrial ablation procedures at our Institution. As our approach provides a respiratory motion model of the heart displacement, it may potentially be exploited to obtain motion compensated road maps to guide intervention as well. In addition, the feasibility of this framework will be tested after contrast administration. The original BOOST implementation exploiting T2Prep28 has been already successfully applied to a patient cohort in combination with gadolinium injection for simultaneous bright‐blood heart anatomy visualization and black‐blood late gadolinium enhancement (LGE) detection.46 As MT‐preparation has been introduced as an alternative to T2Prep for IR‐prepared black‐blood LGE assessment,47 the framework proposed in this study holds promise for post‐interventional black‐blood LGE characterization of ablation lesions. Furthermore, 3D comprehensive assessment of cardiovascular anatomy in patients with congenital heart disease is gaining importance.48, 49, 50 The proposed BOOST implementation might provide diagnostic insights in such populations as well. Additionally, the assessment of the subject‐specific anatomy of the CS and of the coronary vein morphology is critical in the planning of procedures for cardiac resynchronization therapy in patients suffering from heart failure,51, 52 therefore, the use of the MT‐prepared BOOST sequence will be also investigated in this context.

5. CONCLUSION

This study introduces a novel free‐breathing multi‐contrast 3D whole‐heart sequence (MT‐prepared BOOST) that is suitable for non‐contrast enhanced interventional planning of ablation procedures for the treatment of AF. The proposed framework provides simultaneous bright‐blood visualization of the coronary veins and arteries and black‐blood depiction of atrial walls. With this sequence, data acquisition is performed with nearly 100% scan efficiency and with predictable examination time. Non‐rigid respiratory motion correction further improves image quality compared to translational correction alone. Further studies in patients scheduled for atrial ablation are now warranted.

CONFLICT OF INTEREST

Radhouene Neji and Peter Montney are employed by Sienmens Healthcare.

Supporting information

FIGURE S1 Bright‐blood non‐contrast enhanced MTC‐IR BOOST (A, C, E, and G) and bright‐blood contrast enhanced clinical reference (B, D, F, and H) in 2 patients scheduled for an ablation procedure. Images are here presented in coronal (A‐D) and transversal (E‐H) orientations for both sequences.

ACKNOWLEDGMENTS

This work was supported by the following grants: EPSRC EP/N009258/1, EP/P001009/1, EP/P007619/1, MRC MR/L009676/1, and FONDECYT 1161051; the Welcome EPSRC Centre for Medical Engineering (WT203148/Z/16/Z); and the Department of Health via the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award to Guy’s & St. Thomas NHS Foundation Trust in partnership with King’s College London and King’s College Hospital NHS Foundation Trust. The views expressed are those of the authors and not necessarily those of the National Health Service, the National Institute for Health Research, or the Department of Health. Concepts and information presented are based on research and are not commercially available. Because of regulatory reasons, the future availability cannot be guaranteed.

Ginami G, Lòpez K, Mukherjee RK, et al. Non‐contrast enhanced simultaneous 3D whole‐heart bright‐blood pulmonary veins visualization and black‐blood quantification of atrial wall thickness. Magn Reson Med. 2019;81:1066–1079. 10.1002/mrm.27472

Funding information

EPSRC, Grant/Award Numbers: EP/N009258/1, EP/P001009/1, EP/P007619/1, MRC MR/L009676/1; FONDECYT, Grant/Award Numbers: 1161051; Welcome EPSRC Centre for Medical Engineering, Grant/Award Numbers: WT203148/Z/16/Z; National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award to Guy’s; St. Thomas’ NHS Foundation Trust in partnership with King’s College London; King’s College Hospital NHS Foundation Trust

REFERENCES

  • 1. Camm AJ, Lip GY, De Caterina R, et al. 2012 focused update of the ESC guidelines for the management of atrial fibrillation: an update of the 2010 ESC guidelines for the management of atrial fibrillation–developed with the special contribution of the European Heart Rhythm Association. Eur Heart J. 2012;14:1385‐1413. [DOI] [PubMed] [Google Scholar]
  • 2. Ames A, Stevenson WG. Cardiology patient page. Catheter ablation of atrial fibrillation. Circulation. 2006;113:e666‐e668. [DOI] [PubMed] [Google Scholar]
  • 3. Syed MA, Peters DC, Rashid H, Arai AE. Pulmonary vein imaging: comparison of 3D magnetic resonance angiography with 2D cine MRI for characterizing anatomy and size. J Cardiovasc Magn Reson. 2005;7:355‐360. [DOI] [PubMed] [Google Scholar]
  • 4. Mansour M, Holmvang G, Sosnovik D, et al. Assessment of pulmonary vein anatomic variability by magnetic resonance imaging: implications for catheter ablation techniques for atrial fibrillation. J Cardiovasc Electrophysiol. 2004;15:387‐393. [DOI] [PubMed] [Google Scholar]
  • 5. Wittkampf FH, Vonken EJ, Derksen R, et al. Pulmonary vein ostium geometry: analysis by magnetic resonance angiography. Circulation. 2003;107:21‐23. [DOI] [PubMed] [Google Scholar]
  • 6. Kato R, Lickfett L, Meininger G, et al. Pulmonary vein anatomy in patients undergoing catheter ablation of atrial fibrillation: lessons learned by use of magnetic resonance imaging. Circulation. 2003;107:2004‐2010. [DOI] [PubMed] [Google Scholar]
  • 7. Roujol S, Foppa M, Basha TA, et al. Accelerated free breathing ECG triggered contrast enhanced pulmonary vein magnetic resonance angiography using compressed sensing. J Cardiovasc Magn Reson. 2014;16:91. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Hu P, Chuang ML, Kissinger KV, et al. Non‐contrast‐enhanced pulmonary vein MRI with a spatially selective slab inversion preparation sequence. Magn Reson Med. 2010;63:530‐536. [DOI] [PubMed] [Google Scholar]
  • 9. Akcakaya M, Hu P, Chuang ML, et al. Accelerated noncontrast‐enhanced pulmonary vein MRA with distributed compressed sensing. J Magn Reson Imaging. 2011;33:1248‐1255. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Nezafat R, Han Y, Peters DC, et al. Coronary magnetic resonance vein imaging: imaging contrast, sequence, and timing. Magn Reson Med. 2007;58:1196‐1206. [DOI] [PubMed] [Google Scholar]
  • 11. Stoeck CT, Han Y, Peters DC, et al. Whole heart magnetization‐prepared steady‐state free precession coronary vein MRI. J Magn Reson Imaging. 2009;29:1293‐1299. [DOI] [PubMed] [Google Scholar]
  • 12. Stoeck CT, Hu P, Peters DC, et al. Optimization of on‐resonant magnetization transfer contrast in coronary vein MRI. Magn Reson Med. 2010;64:1849‐1854. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Keegan J, Drivas P, Firmin DN. Navigator artifact reduction in three‐dimensional late gadolinium enhancement imaging of the atria. Magn Reson Med. 2014;72:779‐785. [DOI] [PubMed] [Google Scholar]
  • 14. Moghari MH, Peters DC, Smink J, et al. Pulmonary vein inflow artifact reduction for free‐breathing left atrium late gadolinium enhancement. Magn Reson Med. 2011;66:180‐186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Calkins H, Kuck KH, Cappato R, et al. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow‐up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) task force on catheter and surgical ablation of atrial fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012;9:632‐696. [DOI] [PubMed] [Google Scholar]
  • 16. Whitaker J, Rajani R, Chubb H, et al. The role of myocardial wall thickness in atrial arrhythmogenesis. Europace. 2016;18:1758‐1772. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Cappato R, Calkins H, Chen SA, et al. Prevalence and causes of fatal outcome in catheter ablation of atrial fibrillation. J Am Coll Cardiol. 2009;53:1798‐1803. [DOI] [PubMed] [Google Scholar]
  • 18. Bunch TJ, Asirvatham SJ, Friedman PA, et al. Outcomes after cardiac perforation during radiofrequency ablation of the atrium. J Cardiovasc Electrophysiol. 2005;16:1172‐1179. [DOI] [PubMed] [Google Scholar]
  • 19. Yamazaki M, Mironov S, Taravant C, et al. Heterogeneous atrial wall thickness and stretch promote scroll waves anchoring during atrial fibrillation. Cardiovasc Res. 2012;94:48‐57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Beinart R, Abbara S, Blum A, et al. Left atrial wall thickness variability measured by CT scans in patients undergoing pulmonary vein isolation. J Cardiovascular Electrophysiol. 2011;22:1232‐1236. [DOI] [PubMed] [Google Scholar]
  • 21. Nakamura K, Funabashi N, Uehara M, et al. Left atrial wall thickness in paroxysmal atrial fibrillation by multislice‐CT is initial marker of structural remodeling and predictor of transition from paroxysmal to chronic form. Int J Cardiol. 2011;148:139‐147. [DOI] [PubMed] [Google Scholar]
  • 22. Hoffmeister PS, Chaudhry GM, Mendel J, et al. Evaluation of left atrial and posterior mediastinal anatomy by multidetector helical computed tomography imaging: relevance to ablation. J Interv Card Electrophysiol. 2007;18:217‐223. [DOI] [PubMed] [Google Scholar]
  • 23. Bishop M, Rajani R, Plank G, et al. Three‐dimensional atrial wall thickness maps to inform catheter ablation procedures for atrial fibrillation. Europace. 2016;18:376‐383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Varela M, Morgan R, Theron A, et al. Novel MRI technique enables non‐invasive measurement of atrial wall thickness. IEEE Trans Med Imaging. 2017;36:1607‐1614. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Kellman P, Arai AE, McVeigh ER, Aletras AH. Phase‐sensitive inversion recovery for detecting myocardial infarction using gadolinium‐delayed hyperenhancement. Magn Reson Med. 2002;47:372‐383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole‐heart coronary MR angiography with 2D self‐navigated image reconstruction. Magn Reson Med. 2012;67:437‐445. [DOI] [PubMed] [Google Scholar]
  • 27. Cruz G, Atkinson D, Henningsson M, Botnar RM, Prieto C. Highly efficient nonrigid motion‐corrected 3D whole‐heart coronary vessel wall imaging. Magn Reson Med. 2017;77:1894‐1908. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Ginami G, Neji R, Phinikaridou A, Whitaker J, Botnar RM, Prieto C. Simultaneous bright‐ and black‐blood whole‐heart MRI for noncontrast enhanced coronary lumen and thrombus visualization. Magn Reson Med. 2018;79:1460‐1472. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Prieto C, Doneva M, Usman M, et al. Highly efficient respiratory motion compensated free‐breathing coronary MRA using golden‐step Cartesian acquisition. J Magn Reson Imaging. 2015;41:738‐746. [DOI] [PubMed] [Google Scholar]
  • 30. Atlas SW, Grossman RI, Hackney DB, Goldberg HI, Bilaniuk LT, Zimmerman RA. STIR MR imaging of the orbit. AJR Am J Roentgenol. 1988;151:1025‐1030. [DOI] [PubMed] [Google Scholar]
  • 31. Haase A, Frahm J, Hanicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30:341‐344. [DOI] [PubMed] [Google Scholar]
  • 32. Sussman MS, Wright GA. Factors affecting the correlation coefficient template matching algorithm with application to real‐time 2‐D coronary artery MR imaging. IEEE Trans Med Imaging. 2003;22:206‐216. [DOI] [PubMed] [Google Scholar]
  • 33. Piccini D, Bonanno G, Ginami G, Littmann A, Zenge MO, Stuber M. Is there an optimal respiratory reference position for self‐navigated whole‐heart coronary MR angiography? J Magn Reson Imaging. 2016;43:426‐433. [DOI] [PubMed] [Google Scholar]
  • 34. Pruessmann KP, Weiger M, Bornert P, Boesiger P. Advances in sensitivity encoding with arbitrary k‐space trajectories. Magn Reson Med. 2001;46:638‐651. [DOI] [PubMed] [Google Scholar]
  • 35. Forman C, Piccini D, Grimm R, Hutter J, Hornegger J, Zenge MO. Reduction of respiratory motion artifacts for free‐breathing whole‐heart coronary MRA by weighted iterative reconstruction. Magn Reson Med. 2015;73:1885‐1895. [DOI] [PubMed] [Google Scholar]
  • 36. Correia T, Cruz G, Schneider T, Botnar RM, Prieto C. Accelerated nonrigid motion compensated isotropic 3D coronary MR angiography. Med Phys. 2018;45:214‐222. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Modat M, Ridgway GR, Taylor ZA, et al. Fast free‐form deformation using graphics processing units. Comput Methods Programs Biomed. 2010;98:278‐284. [DOI] [PubMed] [Google Scholar]
  • 38. Batchelor PG, Atkinson D, Irarrazaval P, Hill DL, Hajnal J, Larkman D. Matrix description of general motion correction applied to multishot images. Magn Reson Med. 2005;54:1273‐1280. [DOI] [PubMed] [Google Scholar]
  • 39. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
  • 40. Etienne A, Botnar RM, Van Muiswinkel AM, Boesiger P, Manning WJ, Stuber M. "Soap‐Bubble" visualization and quantitative analysis of 3D coronary magnetic resonance angiograms. Magn Reson Med. 2002;48:658‐666. [DOI] [PubMed] [Google Scholar]
  • 41. Smith TC, Frank E. Introducing machine learning concepts with WEKA. Methods Mol Biol. 2016;1418:353‐378. [DOI] [PubMed] [Google Scholar]
  • 42. Schindelin J, Arganda‐Carreras I, Frise E, et al. Fiji: an open‐source platform for biological‐image analysis. Nat Methods. 2012;9:676‐682. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Ginami G, Bonanno G, Schwitter J, Stuber M, Piccini D. An iterative approach to respiratory self‐navigated whole‐heart coronary MRA significantly improves image quality in a preliminary patient study. Magn Reson Med. 2016;75:1594‐1604. [DOI] [PubMed] [Google Scholar]
  • 44. Munoz C, Neji R, Cruz G, et al. Motion‐corrected simultaneous cardiac positron emission tomography and coronary MR angiography with high acquisition efficiency. Magn Reson Med. 2018;79:339‐350. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Haissaguerre M, Jais P, Shah DC, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339:659‐666. [DOI] [PubMed] [Google Scholar]
  • 46. Ginami G, Neji R, Rashid I, et al. 3D whole‐heart phase sensitive inversion recovery CMR for simultaneous black‐blood late gadolinium enhancement and bright‐blood coronary CMR angiography. J Cardiovasc Magn Reson. 2017;19:94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Kim HW, Rehwald WG, Jenista ER, et al. Dark‐blood delayed enhancement cardiac magnetic resonance of myocardial infarction. JACC Cardiovasc Imaging. 2017. 10.1016/j.jcmg.2017.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Monney P, Piccini D, Rutz T, et al. Single centre experience of the application of self navigated 3D whole heart cardiovascular magnetic resonance for the assessment of cardiac anatomy in congenital heart disease. J Cardiovasc Magn Reson. 2015;17:55. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Henningsson M, Hussain T, Vieira MS, et al. Whole‐heart coronary MR angiography using image‐based navigation for the detection of coronary anomalies in adult patients with congenital heart disease. J Magn Reson Imaging. 2016;43:947‐955. [DOI] [PubMed] [Google Scholar]
  • 50. Velasco Forte MN, Byrne N, et al. Interventional correction of sinus venosus atrial septal defect and partial anomalous pulmonary venous drainage: procedural planning using 3D printed models. JACC Cardiovasc Imaging. 2018;11:275‐278. [DOI] [PubMed] [Google Scholar]
  • 51. de Paola AA, Melo WD, Tavora MZ, Martinez EE. Angiographic and electrophysiological substrates for ventricular tachycardia mapping through the coronary veins. Heart. 1998;79:59‐63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Singh JP, Houser S, Heist EK, Ruskin JN. The coronary venous anatomy: a segmental approach to aid cardiac resynchronization therapy. J Am Coll Cardiol. 2005;46:68‐74. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

FIGURE S1 Bright‐blood non‐contrast enhanced MTC‐IR BOOST (A, C, E, and G) and bright‐blood contrast enhanced clinical reference (B, D, F, and H) in 2 patients scheduled for an ablation procedure. Images are here presented in coronal (A‐D) and transversal (E‐H) orientations for both sequences.


Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES