Table 2.
Comparison of glucose values generated using FreeStyle Libre Pro and iPro2 with closest values generated using self‐monitoring blood glucose
n | SMBG (mg/dL) | FSL‐Pro (mg/dL) | iPro2 (mg/dL) | |
---|---|---|---|---|
Overall | 323 | 189.0 (152.0–248.0) | 184.0 (144.0–249.0)* , ** | 195.0 (151.0–247.0) |
Hypoglycemia | 2 | 63.5 | 67.5 | 105.5 |
Euglycemia | 141 | 145.0 (123.5–165.5) | 139.0 (115.0–158.5)* , ** | 148.0 (125.0–167.0)*** |
Hyperglycemia | 180 | 238.0 (207.3–278.8) | 239.5 (201.0–279.0) | 238.0 (206.0–272.0) |
Glycemia is classified based on the glucose values determined by self‐monitoring blood glucose (SMBG). Data are expressed as medians (interquartile range). *P < 0.01 versus SMBG. **P < 0.01 versus iPro2. ***P < 0.01 versus SMBG. Values were compared using Wilcoxon signed‐rank test with Bonferroni correction. FSL‐Pro, FreeStyle Libre Pro.