Table 2.

Comparison of glucose values generated using FreeStyle Libre Pro and iPro2 with closest values generated using self‐monitoring blood glucose

n SMBG (mg/dL) FSL‐Pro (mg/dL) iPro2 (mg/dL)
Overall 323 189.0 (152.0–248.0) 184.0 (144.0–249.0)* , ** 195.0 (151.0–247.0)
Hypoglycemia 2 63.5 67.5 105.5
Euglycemia 141 145.0 (123.5–165.5) 139.0 (115.0–158.5)* , ** 148.0 (125.0–167.0)***
Hyperglycemia 180 238.0 (207.3–278.8) 239.5 (201.0–279.0) 238.0 (206.0–272.0)

Glycemia is classified based on the glucose values determined by self‐monitoring blood glucose (SMBG). Data are expressed as medians (interquartile range). *< 0.01 versus SMBG. **< 0.01 versus iPro2. ***< 0.01 versus SMBG. Values were compared using Wilcoxon signed‐rank test with Bonferroni correction. FSL‐Pro, FreeStyle Libre Pro.