. 2018 Nov 29;10(3):851–856. doi: 10.1111/jdi.12970

Table 4.

Comparison of glucose values between FreeStyle Libre Pro and iPro2

n FSL‐Pro (mg/dL) iPro2 (mg/dL)
Overall 5,555 177.0 (133.0–228.0)* 183.0 (145.0–230.0)
Hypoglycemia 16 74.5 (60.8–76.8)* 47.0 (40.5–65.0)
Euglycemia 2,669 132.0 (108.0–154.5)* 144.0 (120.0–162.0)
Hyperglycemia 2,870 225.0 (196.0–266.0)* 228.0 (202.0–265.0)

Glycemia is classified based on the glucose values determined by the iPro2. Data are expressed as medians (interquartile range). *< 0.01 versus iPro2. Wilcoxon signed‐rank test was used for a comparison. FSL‐Pro, FreeStyle Libre Pro.