TABLE 3.

Group 2 real-time PCR results using the laboratory-developed real-time assay for the detection of Dientamoeba fragilis following implementation of fluorescence thresholds or CT cutoff values

Assay Platform No. positive No. negative % positive CT range
EasyScreen (10) Bio-Rad CFX96 0 200 0 0–0
Laboratory-developed real-time PCR assay (9) Cepheid SmartCycler II (CT cutoff, 37) 5 195 2.5 32.00–45.16
Roche LightCycler 480a 4 196 2.0 39.59–45.00
Bio-Rad CFX96 (threshold, 12,000 RFU × 103) 2 198 1.0 46.93–47.61
ABI 7500 (CT cutoff, 38) 4 194 3.0 32.34–40.11
a

True-positive and false-positive results could not be distinguished via CT cutoff values or fluorescence thresholds on these platforms. (True positives were established via amplicon sequencing.)