Table 2.

Summary of disagreements in critical range and values outside of acceptable comparative range between laboratory and POCT by analyte and manufacturer

Critical Range Disagreement
Analyte Lab to i-STAT Lab to epoc epoc to i-STAT Total
Sodium 1/101 1/102 2/101 4/304 (1.3%)
Potassium 2/101 0/101 1/100 3/302 (1.0%)
Chloride n/a n/a n/a n/a
Creatinine n/a n/a n/a n/a
Hematocrit n/a n/a n/a n/a
Hemoglobin 0/83 0/84 0/100 0/267 (0.0%)
Glucose
 < 5 mmol/L 0/8 0/8 0/8 0/24 (0.0%)
 ≥ 5 mmol/L 0/29 0/29 0/92 0/150 (0.0%)
 Total 3/322 (0.9%) 1/324 (0.3%) 3/401 (0.7%) 7/1047 (0.7%)
p = 0.323*
Outside of the Laboratory to POCT Acceptable Comparative Range
Analyte Lab to i-STAT Lab to epoc epoc to i-STAT Total
Sodium 2/101 2/102 1/101 5/304 (1.6%)
Potassium 10/101 9/101 1/100 20/302 (6.6%)
Chloride 5/101 14/101 10/100 29/302 (9.6%)
Creatinine 4/100 10/98 7/98 21/296 (7.1%)
Hematocrit 0/83 17/84 3/100 20/267 (7.5%)
Hemoglobin n/a n/a n/a n/a
Glucose
 < 5 mmol/L 5/8 2/8 3/8 10/24 (41.7%)
 ≥ 5 mmol/L 6/29 2/29 13/92 21/150 (14.0%)
 Total 32/523 (6.1%) 56/523 (10.7%) 38/599 (6.3%) 126/1645 (7.7%)
p = 0.007*

Note: POCT Point of care testing device

*Two sample test of proportions between i-STAT® compared with laboratory, and epoc® compared with laboratory