. 2019 May 2;19:30. doi: 10.1186/s12873-019-0243-4

Table 3.

Summary of out-of-range and in-range results for i-STAT® and epoc® by individual device and by manufacturer

i-STAT®
Device Out-of-rangea In-rangea Proportion Out-of-range by device
CP1 0 (0.0%) 13 (100.0%) 0 (0.0%)
CP2 4 (23.5%) 13 (76.5%) 4 (14.8%)
CP3 4 (23.5%) 13 (76.5%) 4 (14.8%)
CP4 2 (20.0%) 8 (80.0%) 2 (7.4%)
CP5 2 (18.1%) 9 (81.8%) 2 (7.4%)
CP6 15 (41.7%) 21 (58.3%) 15 (55.5%)
Total Out-of-range 27/104 (26.0%) 77/104 (74.0%) 27 (100.0%)
epoc®
 CP1 5 (38.5%) 8 (61.5%) 5 (10.2%)
 CP2 6 (31.6%) 13 (68.4%) 6 (12.2%)
 CP3 10 (58.8%) 7 (41.2%) 10 (20.4%)
 CP4 6 (54.6%) 5 (45.5%) 6 (12.2%)
 CP5 3 (27.3%) 8 (72.7%) 3 (6.1%)
 CP6 19 (54.3%) 16 (45.7%) 19 (38.8%)
 Total Out-of-range 49/106 (46.2%) 57/106 (53.8%) 49 (100.0%)

aOut-of-range refers to outside of the a priori laboratory to POCT acceptable Comparative Range and vice versa