Skip to main content
. 2019 Mar 14;46(1):63–69. doi: 10.1002/jmd2.12018

Table 1.

LC‐MS/MS settings and conditions

Liquid chromatography gradient conditions
Time (min) Flow rate (mL/min) Solvent A%a Solvent B%a
0 0.15 70 30
2 0.15 30 70
6.5 0.15 30 70
7.5 0.15 70 30
10 0.15 70 30
Mass spectrometer settings
Curtain gas 15 psi
Ion spray voltage −4500 V
Dissolvation temperature 600°C
GS1 and GS2 20 and 40 psi, respectively
Declustering potential −110 V
Collision energy (CE) for different MRM transitions
Succinyl‐CoA 866.2 > 408 (CE −55 eV)
D4‐Succinyl‐CoA 870.2 > 408 (CE −55 eV)
Malonyl‐CoA 852.2 > 808 (CE −40 eV)
a

Solvent A, 10 mM ammonium formate in acetonitrile/water (90/10, vol/vol); Solvent B, 10 mM ammonium formate in acetonitrile/water (20/80, vol/vol).