Skip to main content
. Author manuscript; available in PMC: 2020 Mar 20.
Published in final edited form as: J Pharm Biomed Anal. 2018 Dec 19;166:105–112. doi: 10.1016/j.jpba.2018.12.028

Table 4:

qNMR Statistical Analysisa (mg of oxytocin per vial)

Source Variance Standard deviation (SD) Relative standard deviation (%RSD)
Reproducibility (lab to lab) 0.0072 0.0851 4.73%
Repeatability (sample to sample) 0.0011 0.0332 1.85%
a

The mean for all six labs was 1.798 mg with 95% confidence interval: 1.707 mg to 1.888 mg. None of the labs’ results were excluded.