Table 8:

HPLC Statistical Analysisa (mg of oxytocin per vial)

Source Variance SD RSD
Reproducibility (lab to lab) 0.00168 0.0410 2.3%
Repeatability (sample to sample) 0.00015 0.0123 0.7%
a

The mean value for the six labs was 1.807 mg with 95% confidence interval: 1.769 mg to 1.845 mg.