Table 2. Accuracy of wearable devices in measuring baseline and induced SVT heart rate.
Clinical Situation | Device | ||||
---|---|---|---|---|---|
Apple | Galaxy | Fitbit | p value | ||
No. of patients | 34 | 34 | 33 | ||
Baseline (within ±5 bpm) | 34 (100) | 34 (100) | 31 (94) | * | |
No. of patients with SVT | 28 | 29 | 30 | ||
SVT (within ±5 bpm) | 25 (89) | 26 (90) | 25 (83) | 0.780 | |
SVT (within ±10 bpm) | 28 (100) | 26 (90) | 26 (87) | 0.568 |
Values are presented as number of patients (%).
bpm = beats per minute; SVT = supraventricular tachycardia.
*Statistical analysis not applicable.