Table 3. Results of the atrial and ventricular pacing study.

Pacing cycle length Pacing site and device
Atrial pacing Ventricular pacing
Apple Galaxy Fitbit p value Apple Galaxy Fitbit p value
600 ms (100 bpm)
Attempted 33 30 29 29 25 25
Within ±5 bpm 33 (100) 29 (97) 27 (93) 0.198 29 (100) 25 (100) 20 (80) 0.007
Within ±10 bpm 33 (100) 29 (97) 27 (93) 0.198 29 (100) 25 (100) 22 (88) 0.010
500 ms (120 bpm)
Attempted 32 29 28 32 29 27
Within ±5 bpm 31 (97) 26 (90) 18 (64) 0.000 32 (100) 27 (93) 20 (74) 0.002
Within ±10 bpm 31 (97) 28 (97) 20 (71) 0.062 32 (100) 27 (93) 23 (85) 0.025
430 ms (140 bpm)
Attempted 29 22 24 33 29 29
Within ±5 bpm 26 (90) 20 (91) 19 (79) 0.005 33 (100) 28 (97) 18 (62) <0.001
Within ±10 bpm 26 (90) 20 (91) 21 (88) 0.032 33 (100) 28 (97) 20 (69) 0.005
380 ms (160 bpm)
Attempted 19 16 17 31 29 27
Within ±5 bpm 18 (95) 14 (88) 12 (71) 0.016 31 (100) 24 (83) 10 (37) <0.001
Within ±10 bpm 18 (95) 14 (88) 12 (71) 0.045 31 (100) 24 (83) 13 (48) <0.001
330 ms (180 bpm)
Attempted 9 10 13 31 28 26
Within ±5 bpm 7 (78) 8 (80) 6 (46) * 28 (90) 20 (71) 7 (27) <0.001
Within ±10 bpm 8 (89) 8 (80) 6 (46) 28 (90) 20 (71) 8 (31) <0.001
300 ms (200 bpm)
Attempted 7 7 8 31 27 24
Within ±5 bpm 5 (71) 3 (43) 3 (38) * 23 (74) 20 (74) 6 (25) 0.002
Within ±10 bpm 5 (71) 3 (43) 3 (38) 23 (74) 20 (74) 7 (29) 0.005

Values are presented as number of patients (%).

*Statistical analysis not applicable.