Table 2. Correlation between LIC and VIC in the model of liver-iron-fat under different tube currents (x¯±s).

Concentrations of fat (%) Tube currents (mA) Samples Spearman analysis Linear correlation equation Adjusted R2 value P value F value
Correlation coefficient P value
0 (Iron-water) 200 6 1.000 0.000    y=1.076x−1.602 0.995 0 994.169
320 6 1.000 0.000    y=1.104x−1.193 0.995 0 969.015
284 6 1.000 0.000    y=1.089x−1.461 0.996 0 1,112.086
0 (Iron-fat) 200 6 1.000 0.000    y=1.083x−10.756 0.993 0.000 662.910
320 6 1.000 0.000    y=1.095x−10.074 0.993 0.000 750.433
485 6 1.000 0.000    y=1.100x−10.621 0.993 0.000 733.509
2.5 200 6 1.000 0.000    y=1.823x−15.546 0.972 0.000 176.750
320 6 1.000 0.000    y=1.924x−15.869 0.962 0.000 126.074
485 6 1.000 0.000    y=1.894x−16.093 0.950 0.001 96.823
7.5 200 6 1.000 0.000    y=1.582x−12.708 0.950 0.001 96.823
320 6 1.000 0.000    y=1.661x−12.643 0.968 0.000 152.418
485 6 1.000 0.000    y=1.721x−13.440 0.960 0.000 120.547
15 200 5 1.000 0.037    y=1.189x−7.078 0.813 0.036 27.489
320 5 1.000 0.000    y=1.603x−10.238 0.936 0.005 59.715
485 5 1.000 0.000    y=1.488x−9.613 0.910 0.008 41.455

LIC, liver iron content; VIC, virtual iron content.