TABLE 1. American Thoracic Society/European Respiratory Society spirometry quality criteria [8].
Acceptability |
---|
Free of artifacts |
Cough |
Glottis closure |
Early termination |
Submaximal effort |
Leaks |
Obstructed mouthpieces |
Back extrapolated volume <5% of FVC or 0.15 L, whichever is greater |
Expiratory time ≥ 6 s, 3 s for children or plateau in the volume-time curve or when the patient cannot or should not continue to exhale |
Repeatability |
A minimum of three acceptable tests should be obtained |
The two largest FVC should be within .15 L of each other |
The two largest FEV1 should be within .15 L of each other |
Note: FVC, forced vital capacity; FEV1 , forced expiratory volume in the first second.