Table 3.

Comparison among different DENV detection assays for 153 clinical samples.

Methods NS1 antigen assay RT-PCR The conventional RT-LAMP Total Positive rate (%)

Items Pos. Neg. Concordance rate (%) Pos. Neg. Concordance rate (%) Pos. Neg. Concordance rate (%)
The novel mismatch- Pos. 136 9 90.8% 116 29 78.4% 133 12 92.2% 145 94.8%
tolerant RT-LAMP Neg. 5 3 4 4 0 8 8
Total 141 12 NA 120 33 NA 133 20 NA 153 NA
Positive rate (%) 92.2% NA 78.4% NA 86.9% NA NA NA

The concordance rate was calculated using the formula: (number of consistent results by both methods/total number) × 100%. Please also see the Venn diagram in Supplementary Figure S3 for details. Pos., positive; Neg., negative; NA, not applicable.