Skip to main content
. Author manuscript; available in PMC: 2019 May 15.
Published in final edited form as: Metabolomics. 2017 Dec 23;14(1):15. doi: 10.1007/s11306-017-1310-z

Table 2.

The r2 and limit of quantification (LOQ: signal to noise ratio > 10) of each analyte quantified using this LC/MS/ MS assay

Analyte Internal Standard Used LOQ (pmol on
column)
r2 (Stds in buffer) aMatrix
effect
(%)

NAM 2H4-NAM 0.2 0.9909 + 10
Adenosine 13C5-ADO 0.05 0.9934  0
MeNAM 2H4-NAM 0.01 0.9918 –20
NA 2H4-NAM 2.0 0.9949 +17
cAMP 13C5-Cyclic AMP 0.05 0.9985 –4
NMN 13C5-Cyclic AMP 0.1 0.9894 –4
NAD+ 13C5-Cyclic AMP 0.1 0.9913 –3
NAAD 2H4-NAM 0.02 0.9882 –18
NADH 13C1015N5-ATP 0.2 0.9878 –17
AMP 13C5-AMP 0.2 0.9912 –4
cADPR 2H4-AMP 0.2 0.9929  0
ADPR 2H4-AMP 0.2 0.9911  0
NAMN 2H4-AMP 0.2 0.9821  –20
NADP+ 15N5-ADP 0.04 0.9901 +7
ADP 15N5-ADP 0.4 0.9958 +6
NADPH 2H4-NAM 0.2 0.9923 –7
ATP 13C1015N5-ATP 0.4 0.9929 –5

The standard curves are shown in the supplementary file 1

a

positive values denote signal enhancement and negative values indicate signal suppression