Skip to main content
. 2019 Jun 15;1118-1119:33–39. doi: 10.1016/j.jchromb.2019.04.026

Table 1.

Measured HPLC peak area for different PK11195 concentrations, repeatability and reproducibility of the developed analytical method. All data shown as average response±SD (n = 6), D1 = the 1st day of sample analysis and D2 = the 2nd day of sample analysis.

Concentration of PK11195
7.5 ng/mL 10 ng/mL 15 ng/mL 25 ng/mL
Average response D1 (mAU·min) 0.010 ± 0.002 0.011 ± 0.001 0.012 ± 0.002 0.020 ± 0.003
Repeatability D1 (%) 22 11 7 5
Average response D2 (mAU·min) 0.009 ± 0.001 0.010 ± 0.001 0.013 ± 0.001 0.023 ± 0.001
Repeatability D2 (%) 14 13 5 6
Intermediate precision D1 vs. D2 (%) 24 13 15 17