. 2019 Apr 11;20(5):84–98. doi: 10.1002/acm2.12583

Table 3.

Quantitative comparison between planned and measured dose distributions of non‐coplanar arcs delivered to the block phantom. Nine 160° arcs at six different couch angles were planned to deliver 8 Gy at isocenter using 10 MV flattening filter free beams. Data are averages of 10 measurements, defined as in the caption of Table 2

Field size (mm2) MU Average dose error (%) Max dose error (%) Average error Δx (mm) Average error Δy (mm) Average vector error v (mm) Maximum vector error v (mm) γ>1 3%/1 mm (%)
2.4 × 5, 9 arcs 2477 0.3 ± 2.0 3.1 0.05 ± 0.13 0.13 ± 0.10 0.20 ± 0.08 0.43 0.14