Skip to main content
. 2019 Apr 1;9(4):502. doi: 10.3390/nano9040502

Table 3.

Uncertainty budget for small angle X-ray scattering (SAXS) for the Au30 sample.

Input xi xi · Unit u(xi) · Unit uC/mL1
s 172.1μm 0.2μm 6.1·109
L 4511mm 5mm 5.7·109
Φ0 6.8·109ph/s 6.8·107ph/s 1.85·109
T 2.00% 0.02% 1.85·109
E 8000.0eV 0.8eV 5.6·108
t 60.00s <0.06s 1.85·108
ηQE 97% 3% 6.0·107
Nfit 2.66·105 1.7·106 9.58·109
w 1.01mm 0.03mm 5.46·109
(Δρ)2 1.88·107nm6 4.05·104nm6 4.25·108
C 1.85·1011ml1 1.3·1010