Skip to main content
. 2019 Apr 8;9(4):69. doi: 10.3390/metabo9040069

Table 1.

Intra-day and inter-day analytical validations of gastrodin and 4-hydroxybenzyl alcohol (4-HBA) in rat plasma.

Spiked Concentrations (µg/mL) Intra-day (n = 5) Inter-day (n = 5)
Accuracy % CV % Accuracy % CV %
Gastrodin 0.01 97.0 ± 10.5 10.8 106.6 ± 13.2 12.4
0.1 94.3 ± 4.5 4.8 93.4 ± 6.1 6.6
1 101.6 ± 4.8 4.7 98.5 ± 3.8 3.9
20 93.5 ± 5.8 6.2 94.6 ± 3.1 3.3
4-HBA 0.1 106.0 ± 7.8 7.4 115.0 ± 2.8 2.5
1 102.7 ± 1.5 1.5 99.0 ± 4.2 4.3
5 105.0 ± 8.3 7.9 94.1 ± 4.4 4.6
10 109.7 ± 8.4 7.6 112.2 ± 3.3 3.0