Published in final edited form as: J Magn Reson Imaging. 2018 Oct 22;49(6):1617–1628. doi: 10.1002/jmri.26539

TABLE 1.

ACRIN 6698 DW-MRI Protocol Parameters and Protocol Adherence for N = 89 Test/Retest Cohort

Parameter DWI protocol specification Test/retest cohort values (N) Nacq out of protocola
Acquisition order DWI precontrast injection Compliant (88), mixed (1)b 1
Sequence type SS-EPI All compliant
Frequency direction R/Lc R/L (60), A/P (10), mixed (1)d
FOV (mm) 260–360 280–390 4
Matrix-frequency 128–192 128–196 2
Matrix-phase 128–192 128–194 2
Fat-suppression Active fat-sat All compliant
TR (msec) ≥4,000 4800–9400
TE (msec) Minimum TE (50–100) 59.2–114.0 5
Flip Angle 90 degrees 90
B values 0, 100, 600, 800 s/mm2 All compliant
Slice thickness (mm) 4–5 4 (44), 5 (27)
Number of slices Complete bilateral coverage 24–50
Slice gap (mm) ≤ 1.0 0
No. averages ≥2 2–5
Parallel imaging factor ≥2 N/C
Total acquisition time ≤ 5 minutes N/C

SS-EPI: Single-shot echo-planar imaging, N/C: data not collected.

a

Number of acquisitions (test or retest) with out-of-protocol values.

b

One study was done with the retest DWI acquisition following the contrast injection.

c

Protocol requirement of R/L frequency encoding was relaxed after investigation indicated better image quality with A/P frequency on some scanners.

d

One study was done with reversed phase/frequency encoding directions between test and retest scans.