Table 3:

Typical Diffusion-Weighted Imaging (DWI) and Dynamic Contrast Enhanced (DCE) Acquisition Details for (A) Brain, (B) Prostate, (C) Breast, (D) Liver, and (E) Head and Neck

(A) BRAIN
Parameters DWI DCE
Field Strength 1.5 T/3T 1.5T/3T
Acquisition Sequence SS-EPI 3D SPGR
Receive Coil type ≥8channel head array coil ≥8channel head array coil
Lipid Suppression On On
Slice thickness (mm) 4–5 ≤5
Gap thickness (mm) 0–2 0–1
FOV(mm) 220 −240 220 −240
Acquisition Matrix 160–256 × 160–256 or 1.5–1 mm in plane resolution 256×128–160
Plane Orientation Axial Axial
Phase/frequency encode direction AP/RL AP/RL
Receiver bandwidth (Hz/pixel) Max possible in freq encoding direction (acceptable > 1000) 250
DWI Sequence Class
  • Monopolar

  • Bipolar

  • Bipolar Double Spin Echo

Contrast* Pre- Post-
# b-values ≥ 2 (including b=0 s/mm2) # Phases ≥5 40–80
Specific Parameters Minimum highest b-value (s/mm2) 850–1000 # Averages ≥1 1
# Averages ≥2 Flip Angles (FAs) (deg) 2–30* 25–30
Diffusion Directions 3 orthogonal # Flip Angles (FAs) 2–7 1
TR (ms) 3000–5000 TR (ms) 3–8¥ 3–8
TE (ms) Minimum TE (ms) ≤3¥ ≤3
In plane parallel imaging 2 Temporal Resolution (sec) /Total Acquisition Time (min) <10 (ideal 5)/5–10
Total Acquisition Time (min) 3 #Contrast Dose and IV injection rate see references
*Variable FAs for T10 measurement
¥Ensure TR/TE stays constant for all flip angles
(B) PROSTATE
Parameters DWT DCE
Field Strength 3T 1.5T/3T
Acquisition Sequence SS-EPI 3D SPGR
Receive Coil type > 8 channel torso array coil; pelvic phase array/endorectal coils; body array coil) > 8 channel torso array coil; pelvic phase array coil/endorectal coils; body array coil)
Lipid Suppression On NA
Slice thickness (mm) 3–5 <5
Gap thickness (mm) 0–1 0–1
FOV(mm) 240–260 260–300
Acquisition Matrix 224–128 × 224–128; 1–2 mm in-plane resolution <256×l60
Plane Orientation Axial Axial
Phase/frequency encode direction AP/RL RL/AP
Receiver bandwidth (Hz/pixel) Max possible in freq encoding direction (acceptable > 1000) 250
DWI Sequence Class
  • Monopolar

  • Bipolar

  • Bipolar Double Spin Echo

Contrast# Pre- Post-
# b-values 2 (including b < 50–100 s/mm2) # Phases ≥5 Sufficient to allow acquisition of at least 5 min post injection; ≥ 30
Specific Parameters Minimum highest b-value (s/mm2) 500–1500 # Averages ≥1 1
# Averages 2-≥4 Flip Angles (FAs) (deg) 2–15* 10–15
Diffusion Directions ≥ 3 orthogonal #Flip Angles
(FAs)
3–5 1
TR (ms) ≤4000 TR (ms) <5¥ <5
TE (ms) Minimum TE (ms) ≤2¥ ≤2
In plane parallel imaging 2 Temporal Resolution (sec) /Total Acquisition Time (min) <10/5–10
Total Acquisition Time (min) 3 #Contrast Dose and IV injection rate see references
* Variable FAs for T10 measurement
¥Ensure TRITE stays constant for all flip angles
(C) BREAST
Parameters DWI DCE
Field Strength 1.5T/3T 1.5T/3T
Acquisition Sequence SS-EPI/SE-EPI 3D SPGR
Receive Coil type ≥ 4 channel breast phase array coil > 4 channel breast phase array coil
Lipid Suppression On On
Slice thickness (mm) 4–5 <2.5
Gap thickness (mm) 0–1 0
FOV(mm) 260–360 To cover the entire breast whether uni- or bilateral
Acquisition Matrix 128–192 × 128–192 > 192×256; 1–1.5 mm in-plane resolution
Plane Orientation Axial Sagittal for single breast coverage; axial for bi lateral coverage
Phase/frequency encode direction RL/AP RL/AP for axial bilateral; HF/AP for sagittal unilateral
Receiver bandwidth (Hz/pixel) Max possible in freq encoding direction 250
DWI Sequence Class
  • Monopolar

  • Bipolar

  • Bipolar Double Spin Echo

Contrast# Pre- Post-
# b-values ≥2 # Phases ≥2 Sufficient to allow acquisition for ≥ 8 min post injection
Specific Parameters Minimum highest b-value (s/mm2) 800 # Averages ≥1 1
# Averages ≥2 Flip Angle (FA) (deg) 2–30* 10–30
Diffusion Directions 3 orthogonal # Flip Angles (FAs) 3–5 1
TR (ms) ≥4000 TR (ms) <8¥ <8
TE (ms) Minimum TE (ms) ≤3¥ ≤3
In plane parallel imaging 2 Temporal Resolution (sec) /Total Acquisition Time (min) ≤ 20 /≥8
Total Acquisition Time (min) 4–6 * Contrast Dose and IV injection rate see references
* Variable FAs for T10 measurement
¥ Ensure TR/TE stays constant for all flip angles
(D) LIVER
Parameters DWI DCE
Field Strength 1.5T/3T 1.5T/3T
Acquisition Sequence SS-EPI 3D SPGR
Receive Coil type > 6–16 channel torso array coil > 8–32 channel flexible or AP body array coil
Lipid Suppression On On
Slice thickness (mm) 5–7 4–5
Gap thickness (mm) 0–1 0
FOV(mm) 300–450 280–380
Acquisition Matrix 160–196 × 160–192 or 2 −3 in-plane resolution 320 × (160–192)
Plane Orientation Axial Oblique
Phase/frequency encode direction AP/RL AP/RL
Receiver bandwidth (Hz/pixel) Max possible in freq encoding direction (acceptable > 1000) 250
DWI Sequence Class
  • Monopolar

  • Bipolar

  • Bipolar Double Spin Echo

Contrast# Pre- Post-
# b-values ≥2 (including one b = 50–100 s/mm2) # Phases ≥5 100
Specific Parameters Minimum highest b-value (s/mm2) 600–800 # Averages ≥2 1
# Averages 2-≥4 Flip Angles (FAs) (deg) 2–30* 20–30
Diffusion Directions 3 orthogonal # Flip Angles (FAs) 3–5 1
TR (ms) >2000 TR (ms) 3–7¥ 3–7
TE (ms) Minimum TE (ms) ≤5¥ ≤5
In plane parallel imaging 2–3 Temporal Resolution (sec) /Total Acquisition Time (min) < 5(ideal~3)/5
Total Acquisition Time (min) ~5 # Contrast Dose and IV injection rate see reference
* Variable FAs for T10 measurement
¥Ensure TR/TE stays constant for all flip angles
(E) HEAD AND NECK
Parameters DWI DCE
Field Strength 1.5T/3T 1.5 T/3T
Acquisition Sequence SS-EPI 3D SPGR
Receive Coil type Neck array or neurovascular coil Neck array or neurovascular coil
Lipid Suppression On On
Slice thickness (mm) ≥5 >5
Gap thickness (mm) 0 0
FOV(mm) 220–380 180–220
Acquisition Matrix 128×128 256×128
Plane Orientation Axial Axial
Phase/frequency encode direction AP/RL RL/AP
Receiver bandwidth (Hz/pixel) Max possible in freq encoding direction 250
DWI Sequence Class
  • Monopolar

  • Bipolar

  • Bipolar Double Spin Echo

Contrast# Pre- Post-
# b-values ≥3 # Phases ≥5 40–80
Specific Parameters Minimum highest b-value (s/mm2) 1000 # Averages ≥ 1 1
# Averages >2 Flip Angles (FAs) (deg) 5–30* 15–30
Diffusion Directions 3 orthogonal # Flip Angles (FAs) ≥3 1
TR (ms) ≥2000 TR (ms) <9¥ <9
TE (ms) Minimum TE (ms) <2¥ <2
In plane parallel imaging 2 Temporal Resolution (sec) /Total Acquisition Time (min) ≤6/~5
Total Acquisition Time (min) 3 #Contrast Dose and IV injection rate see references
*Variable FAs for T10 measurement
¥Ensure TR/TE stays constant for all flip angles