Skip to main content
. 2019 Feb 12;6(1):48–52. doi: 10.14744/nci.2018.44452

TABLE 1.

Average heart rate, QT, corrected QT, and Pearson’s correlation using smartphone electrocardiography and 12-lead electrocardiography

Electrocardiography Pearson’s correlation (r)
Mean heart rate (BPM±SD)
 12-lead electrocardiography 92±20 0.92
 Smartphone electrocardiography 93±22
Mean QT interval (ms±SD)
 12-lead electrocardiography 343±40 0.83
 Smartphone electrocardiography 340±41
Mean corrected QT interval (ms±SD)
 12-lead electrocardiography 419±28 0.57
 Smartphone electrocardiography 415±33

SD: Standard deviation.