Table 2: Steps to Minimise Issues During Deep Brain Stimulation and Cardiovascular Implantable Electronic Device Implantation.

  • Multidisciplinary team approach with neurology input.

  • Implant cardiac device on the contralateral side to the implanted pulse generator.

  • Ensure there is at least 20 cm between the two generators.

  • Use bipolar sensing configuration for the cardiac device.

  • Ideally, use bipolar deep brain stimulation settings with a minimum frequency of 60 Hz.

  • Consider leadless pacemaker in patients with bilateral IPGs.

  • Avoid diathermy during cardiovascular implantable electronic device implantation.

  • Programme devices for ‘worst case scenario’ settings during testing.