TABLE 1.

Values Selected for the Key Imaging Parameters of the Three Sequences

Scan Parameter 2D TSE PD-Weighted Sagittal 2D TSE T2-Weighted Fat Saturated Sagittal 3D GE (FLASH) Sagittal
Repetition time TR, ms 3520 3520 11
Echo time TE, ms 12 63 3.7
Flip angle 153 147 20
Averages 1 1 1
Pixel bandwidth 250 199 250
Echo train length 5 5 1
Width × height, mm 160.0 × 160.0 160.0 × 160.0 160.0 × 160.0
Acquisition matrix (rows × columns) 512 × 512 512 × 512 512 × 512
Slice thickness, mm 1.2 1.2 1.2
Number of slices 36 36 36
Space between slices, mm 1.44 1.44
Acquisition time, min 6.01 6.0 3.23
Resolution (pixels per mm) 3.2 3.2 3.2
Voxel size, mm 0.3125 × 0.3125 × 1.2 0.3125 × 0.3125 × 1.2 0.3125 × 0.3125 × 1.2
Chemical shift, pixels ∼1.8 ∼2.5 ∼1.8

The chemical shift was calculated using the pixel bandwidth (pxBW) and resonance frequency difference between fat and water at 3 T (∼447 Hz). TSE indicates Turbo-Spin-Echo.