Checklist for Transsphenoidal Pituitary Tumor Resections Using the ioMRI Suite

Before sterile draping
Questions to OR/circulation nurse Response
Lower head coil installed?* Water-tightly packed (cover: 80 × 80 cm).
Sticky cloth (75 × 90 cm)a?* Attached to the shoulder.
2 ×Moltexb?* Placed before and behind pins.
PEEK pins installedc?** Quantity:
Earplugs?** Applied bilaterally.
Electrocautery plate installed?* Location:
Hands, Feet?* Positioned straight.
Phantom?* Fits well.
Questions to anesthesiology team Response
Electrical cables?** Not in contact with the patient.
Neck?* Tamponated with fabric.
Tracheal tube?* Positioned flat.
Hose?* Not fixed with towel clamps.
Patient positioned on a dry cover (after insertion of the urinary catheter)?* Double-checked.
After sterile draping
Questions to OR/circulation nurse Response
Sterile towel clamps?** Quantity:
“Spitztupfer” installedd?* Quantity:
Before transfer covering
Questions to OR/circulation nurse Response
Scar?*** “Netzel” implantede
Collection bag fluid below head?* Removed.
Sterile towel clamps?** Removed. Quantity:
Navigation star?** Removed.
Electrocautery plate and cable?** Removed.
After transfer covering
Questions to OR/circulation nurse Response
Upper head coil installed?* Installed. Mind the “little man”! (°)
Cover 125cmf?* Overlaid.
Covering sheets?* Shortened.
White transfer plate? (°)* Visible.
OR table, docking?* Brought into transfer position.
Sliding door MRI?* Free of obstacles.
Coils? Connectors?* Exposed.
Date: __________________
Signature: ___________________ Patient etiquette:

The ** indicate the main rationale: * = organizational; ** = magnetic field strength-related; *** = sterility. Most steps are relevant to any 2-room ioMRI-OR suite, whereas some are specifically relevant for the Siemens magnet (Siemens Healthineers, Erlangen, Germany) with NORAS (Noras MRI products, Höchberg, Germany) head coil (marked with °)

aProduct ID 2626, Mundus Medici GmbH, Düsseldorf, Germany; bProduct ID 161008, MoliNea Plus, Hartmann, Neuhausen, Switzerland; cPEEK pins reusable, NORAS MRI products, Höchberg, Germany; dProduct ID 35003, Novimed Medizintechnik, Dietikon, Switzerland; e Product ID 35004, Novimed Medizintechnik, Dietikon, Switzerland; f Product ID 401008125, Schlieren, Switzerland.