Table 2. Implantable cardioverter-defibrillators – responses to applying magnet.
Manufacturer | Anti-bradycardia function | Anti-tachycardia function |
---|---|---|
Biotronik | None | Detection/therapy - OFF+ |
Boston Scientific# | None | Multiple answers are possible. With some devices, the magnet is applied only once, with others it must be applied continuously, and with some the response to magnet can be turned off |
Medtronic | None | Detection/therapy - OFF |
St. Jude Medical | None | Detection/therapy - OFF* |
Sorin# | Asynchronous mode (AOO, VOO, DOO), frequency 96/min | Detection/therapy - OFF |
#nowadays very rare in Croatia; *response can be switched-off with programming the pacemaker; +in new ICD generation during 8 hours