Table 2. Procedural characteristics and follow-up.
Pulmonary veins isolation (PVI). Mean ±SD, numbers (%), median [25-75 percentiles].
| Procedural characteristics and follow-up | Abbott Safire N=37 | Boston Scientific MiFi N=13 | P value |
|---|---|---|---|
| Ablation: | |||
| Number of RF-applications | 6 [3.5-10] | 8 [2.5-14] | 0.838 |
| Average duration per application (sec) | 70.2 ±33.7 | 60.5 ±22.6 | 0.343 |
| Average temperature (Celsius) | 49.5 ±1.9 | 52.53 ±3.5 | 0.059 |
| Average impedance (ohm) | 74.5 ±5.6 | 74.2 ±6.1 | 0.912 |
| Average power (Watts) | 46.4 ±11.6 | 45.24 ±12.9 | 0.365 |
| Total ablation time (sec) | 456 [263-661.8] | 364 [181-821] | 0.550 |
| Procedure: | |||
| Concomitant PVI | 0 | 1 (7.7) | 0.260 |
| Cross-over | 0 (0) | 0 (0) | - |
| Steerable sheath | 2 (5.4) | 2 (15.4) | 0.275 |
| Time to bi-directional block (min) | 23 [12.5-37.5] | 16 [3-23] | 0.217 |
| Fluoroscopic time (min) | 13 [8-17] | 18 [7-22] | 0.631 |
| Procedure time (min) | 65 [53-80.5] | 50 [34.75-73.25] | 0.133 |
| Follow-up: | |||
| Recurrence of atrial flutter at one-year | 0 (0) | 0(0) | |
| Recurrence of symptoms | 13 (35.1) | 4 (30.8) | 0.774 |
| Occurrence of other atrial arrhythmia | 14 (37.8) | 5 (38.5) | 0.960 |