Table 2. Number and causes of alarm messages.

Patient Type of alarm message Alarms per day
Reduced flow Reduced motor speed Excess current Pump stopped Total
1 1,759 (92.14%) 2 (0.10%) 140 (7.33%) 8* (0.42%) 1,909 4.44
2 1,482 (96.23%) 4 (0.26%) 53 (3.44%) 1 (0.06%) 1,540 3.91
3 27 (44.26%) 2 (3.28%) 31 (50.82%) 1 (1.64%) 61 0.133
4 13 (8.23%) 14 (8.86%) 129 (81.65%) 2 (1.27%) 158 0.486
5 0 9 (47.37%) 9 (47.37%) 1 (5.26%) 19 0.087
6 15 (93.75%) 1 (6.25%) 0 0 16 0.0686
7 59 (67.05%) 2 (2.27%) 27 (30.68%) 0 88 0.727
8 52 (100%) 0 0 0 52 0.481
9 75 (79.79%) 19 (20.21%) 0 0 94 1.28
10 28 (96.55%) 0 1 (3.45%) 0 29 0.557
11 2,244 (99.73%) 6 (0.27%) 0 0 2,250 77.59
Total 5,754 (92.57%) 59 (0.95%) 390 (6.27%) 13 (0.21%) 6,216

*, all 8 of these pump stops were controller-related not pump-related.