Table 2. Number and causes of alarm messages.
Patient | Type of alarm message | Alarms per day | ||||
---|---|---|---|---|---|---|
Reduced flow | Reduced motor speed | Excess current | Pump stopped | Total | ||
1 | 1,759 (92.14%) | 2 (0.10%) | 140 (7.33%) | 8* (0.42%) | 1,909 | 4.44 |
2 | 1,482 (96.23%) | 4 (0.26%) | 53 (3.44%) | 1 (0.06%) | 1,540 | 3.91 |
3 | 27 (44.26%) | 2 (3.28%) | 31 (50.82%) | 1 (1.64%) | 61 | 0.133 |
4 | 13 (8.23%) | 14 (8.86%) | 129 (81.65%) | 2 (1.27%) | 158 | 0.486 |
5 | 0 | 9 (47.37%) | 9 (47.37%) | 1 (5.26%) | 19 | 0.087 |
6 | 15 (93.75%) | 1 (6.25%) | 0 | 0 | 16 | 0.0686 |
7 | 59 (67.05%) | 2 (2.27%) | 27 (30.68%) | 0 | 88 | 0.727 |
8 | 52 (100%) | 0 | 0 | 0 | 52 | 0.481 |
9 | 75 (79.79%) | 19 (20.21%) | 0 | 0 | 94 | 1.28 |
10 | 28 (96.55%) | 0 | 1 (3.45%) | 0 | 29 | 0.557 |
11 | 2,244 (99.73%) | 6 (0.27%) | 0 | 0 | 2,250 | 77.59 |
Total | 5,754 (92.57%) | 59 (0.95%) | 390 (6.27%) | 13 (0.21%) | 6,216 |
*, all 8 of these pump stops were controller-related not pump-related.