Skip to main content
. 2019 May 24;57(6):e00287-19. doi: 10.1128/JCM.00287-19

TABLE 2.

Performance comparison of ELITe InGenius and repMp1 PCRs

Statistical outcome Comparison value (%) (95% confidence interval)
Positive percent agreement 95.79 (89.57–98.84)
Negative percent agreement 99.22 (95.72–99.98)
Positive predictive value 98.91 (92.81–99.84)
Negative predictive value 96.95 (92.40–98.81)
Accuracy 97.76 (94.85–99.27)
Concordance 97.59 (κ = 0.954)
McNemar’s test P = 0.375 (NSa )
a