FIGURE 1.
Parametric images of M0, T1, T2, and T2* from differently prepared specimens. The images were calculated from a series of 2D images (the series sizes reported in subscript of the series range) with resolution 100 μm/pxl × 100 μm/pxl and slice thickness 500 μm. A, FF specimen: T1 (RARE‐VTR), TR (66–3935)9, TE 6; T2 (MSME), TR 3935, TE (6–47)8; T2* (MGE), TR 3935, TE (4–165)24. B, FFPE specimen: T1 (RARE‐VTR), TR (68– 3936)9, TE 9; T2 (RARE‐VTR), TR 3936, TE (9–69)8; T2* (MGE), TR 3936, TE (4.3–83)12. C, Fresh specimen: T1 (RARE‐VTR), TR (110–10000)12, TE 5.8; T2 (RARE‐VTR), TR 10 000, TE (5.8–47)8. D, Gd doped specimen: T1, TR (16–2500)12, TE 5.9; T2, TR 1500, TE (8.4–134.5)16. T2* images are not shown in C because parametric images are grossly distorted and in D because T2* ~ T2